Z-3240-2018 Class II Terminated

Recalled by Stryker GmbH — Selzach

FDA device recall Z-3240-2018 was initiated by Stryker GmbH on Jul 26, 2018 and is designated Class II. Reason for recall: When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously… The recall status is terminated (terminated May 26, 2020). Affected quantity: 233.

Recall Details

Product Type
Devices
Report Date
Oct 3, 2018
Initiation Date
Jul 26, 2018
Termination Date
May 26, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
233

Product Description

Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Reason for Recall

When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.

Distribution Pattern

Nationwide

Code Information

Version #: HLRF 1.0-2/002, HLRF 1.1-2/002

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3240-2018
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026