Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-3286-2018 was initiated by Olympus Corporation Of The Americas on Aug 8, 2018 and is designated Class II. Reason for recall: The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port. The recall status is terminated (terminated May 17, 2022). Affected quantity: 1605.

Recall Details

Product Type
Devices
Report Date
Oct 10, 2018
Initiation Date
Aug 8, 2018
Termination Date
May 17, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1605

Product Description

Bronchovidescope BF-1T240

Reason for Recall

The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.

Distribution Pattern

The products were distributed US nationwide and Canada.

Code Information

all lot codes