Z-3370-2018 Class II Terminated

Recalled by B. Braun Medical, Inc. — Allentown, PA

Recall Details

Product Type
Devices
Report Date
October 10, 2018
Initiation Date
July 20, 2018
Termination Date
May 13, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2400

Product Description

CE17TKPS CONTINUOUS EPIDURAL ANES TRAY, Material Number 332233 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.

Reason for Recall

The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.

Distribution Pattern

Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Guatemala and Kuwait.

Code Information

Batch: 61582595, 61599648, 61605283, 61611124, 61614787