Z-3405-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 10, 2018
- Initiation Date
- July 20, 2018
- Termination Date
- May 13, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 60000
Product Description
BULK NONSTERILE PERIFIX CATHETER CONN., Material Number 8451699N Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.
Reason for Recall
The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.
Distribution Pattern
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Guatemala and Kuwait.
Code Information
Batch: 61606422, 61605072, 61607144, 61618276, 61621554