Stradis Medical, LLC Dba Stradis Healthcare FDA Inspection Record

FEI 1000111025 · Inspections, Form 483s, warning letters and import refusals

Stradis Medical, LLC Dba Stradis Healthcare is an FDA-inspected facility (FEI 1000111025) located in Peachtree Corners, GA, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Aug 5, 2021: No Action Indicated (NAI)
Inspections on record
6 (1 OAI, 2 VAI, 3 NAI)
Last OAI inspection
Apr 12, 2011
Warning letters
1 listed by FDA for this FEI, latest Aug 15, 2011
Recalls as the recalling firm
6, latest Nov 2, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 5, 2021No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Design control - no procedures (21 CFR 820.30(a))
and 10 more
Dec 19, 2016No Action Indicated (NAI)No citations listed
Mar 26, 2014Voluntary Action Indicated (VAI)Process changes - review, evaluation and revalidation (21 CFR 820.75(c))
Production and Process Change Procedures, lack of or Inad. (21 CFR 820.70(b))
Feb 29, 2012No Action Indicated (NAI)Lack of Written MDR Procedures (21 CFR 803.17)
Apr 12, 2011Official Action Indicated (OAI)Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a))
Nonconforming product control (21 CFR 820.90(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 4 more
Apr 3, 2009Voluntary Action Indicated (VAI)Nonconforming product evaluation/investigation (21 CFR 820.90(a))
Training records (21 CFR 820.25(b))
General (21 CFR 820.198(a))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Nov 2, 2022Class IIZ-0481-2023STRADIS HEALTHCARE, WECKCELL EYE SPEARS, Item Numbers: a) 9430, b) 9431 and 77 more products
Medical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely…
Jul 25, 2022Class IIZ-1552-2022NARMD MEDICAL DEPOT Scissors Iris Curved 4.5", Part Number ZZ-0697 and 2 more products
The sterile barrier may be compromised.
Feb 25, 2022Class IIZ-0862-2022Surgicare Neuro Pack - Neurological Kit and 4 more products
The outer bag may be perforated or torn, which may result in a breach in the sterility of the kit.
Oct 28, 2021Class IIZ-0400-2022Custom Assemblies, Inc., Single Spike Tumescent Set w/o Chamber, Reorder Number CMS-1212NS…
This recall has been initiated to address bulk packed, non-sterile Intravascular Administration Sets that were erroneously distributed to…
Sep 10, 2021Class IIZ-0126-2022Tenaculum, Plastic and 5 more products
Firm unable to confirm that the tenaculum devices were produced in accordance with required design and manufacturing standards.
Apr 20, 2015Class IIZ-1992-2015Breast Pack - contains Devon Light Glove Used during surgery and 50 more products
Convenience kits being recalled due to the recall by the manufacturer (Medtronic) of a component (Devon Light Glove) contained in the kits.

FDA record score

9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0481-2023, this FEI as the recalling firmNov 2, 20224 x 0.62.4
Class II recall Z-1552-2022, this FEI as the recalling firmJul 25, 20224 x 0.62.4
Class II recall Z-0862-2022, this FEI as the recalling firmFeb 25, 20224 x 0.62.4
Class II recall Z-0400-2022, this FEI as the recalling firmOct 28, 20214 x 0.251
Class II recall Z-0126-2022, this FEI as the recalling firmSep 10, 20214 x 0.251
Class II recall Z-1992-2015, this FEI as the recalling firmApr 20, 201510 years or older, so it adds no points0
Inspection classified VAIMar 26, 201410 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsAug 15, 201110 years or older, so it adds no points0
Inspection classified OAIApr 12, 201110 years or older, so it adds no points0
Inspection classified VAIApr 3, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Stradis Medical, LLC Dba Stradis Healthcare as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Stradis Medical, LLC Dba Stradis Healthcare yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Stradis Medical, LLC
Location
Peachtree Corners, GA (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.