Trividia Health, Inc. FDA Inspection Record
Trividia Health, Inc. is an FDA-registered facility (FEI 1000113657) located in Fort Lauderdale, FL, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jun 22, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (3 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Feb 6, 2026
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 22, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 1 more |
| Jun 27, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Production processes (21 CFR 820.70(a)) and 1 more |
| Feb 15, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 6, 2026 | Class I | Z-1545-2026 | Owner's Booklets and Instructions for Use that are used with the following blood glucose… and 3 more products The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The… |
| Oct 7, 2025 | Class II | Z-0413-2026 | Good Neighbor Pharmacy, TRUE METRIX Self Monitoring Blood Glucose Meter. Model/Catalog… Manufacturing defect causing LCD display issues with missing or partial segments (ghosting) that can result in misreading glucose values… |
| Apr 20, 2020 | Class II | Z-2201-2020 | TRUE METRIX SELF MONITORING BLOOD GLUCOSE SYSTEM - Product Usage: is intended for the… One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has… |
| Oct 2, 2013 | Class II | Z-0909-2014 | MooreMedical Supporting Health & Care...MooreBrand Blood Glucose Monitor for Self-Testing… and 1 more product Manufactured with an incorrect factory set unit of measure. |
| Feb 25, 2013 | Class II | Z-1140-2013 | Duane Reade TRUEtrack monitor kit, UPC# 6 39194 01679 7 Part # A4006-81
The TRUEtrack… Meter kit outer boxes containing 10 ct. test strips state incorrect storage temperature range. |
FDA record score
22 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-1545-2026, this FEI as the recalling firm | Feb 6, 2026 | 15 x 1 | 15 |
| Class II recall Z-0413-2026, this FEI as the recalling firm | Oct 7, 2025 | 4 x 1 | 4 |
| Inspection classified VAI | Jun 22, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2201-2020, this FEI as the recalling firm | Apr 20, 2020 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jun 27, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0909-2014, this FEI as the recalling firm | Oct 2, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1140-2013, this FEI as the recalling firm | Feb 25, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 15, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 28, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Trividia Health, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Trividia Health, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Trividia Health, Inc. → Check this supplier
Facility details
- Company
- Trividia Health, Inc.
- Location
- Fort Lauderdale, FL (US)
- Source