Pioneer Surgical Technology, Inc. FDA Inspection Record
Pioneer Surgical Technology, Inc. is an FDA-registered facility (FEI 1000115331) located in Marquette, MI, US. RegDataLab tracks 11 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jun 14, 2024: Voluntary Action Indicated (VAI)
- Inspections on record
- 11 (1 OAI, 3 VAI, 7 NAI)
- Last OAI inspection
- Apr 22, 2010
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 3, 2010. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 13, latest Jan 15, 2021
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 14, 2024 | Voluntary Action Indicated (VAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| May 11, 2023 | No Action Indicated (NAI) | No citations listed |
| Mar 28, 2017 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Quality System Procedures (21 CFR 820.20(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 3 more |
| Jun 23, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Sep 24, 2015 | No Action Indicated (NAI) | No citations listed |
| Jul 11, 2014 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) |
| Mar 28, 2012 | No Action Indicated (NAI) | No citations listed |
| Sep 1, 2011 | No Action Indicated (NAI) | No citations listed |
| Dec 21, 2010 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) and 2 more |
| Apr 22, 2010 | Official Action Indicated (OAI) | Promotion or test marketing (21 CFR 812.7(a)) No curriculum vitae in investigator agreement (21 CFR 812.43(c)(1)) Sponsor provide device information to investigators (21 CFR 812.45) Sponsor correspondence records inadequate (21 CFR 812.140(b)(1)) and 6 more |
Warning letters
- Aug 3, 2010: Warning letter listed on FDA's Data Dashboard. View compliance record (114668).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jan 15, 2021 | Class II | Z-1204-2021 | CervAlign Anterior Cervical Plate System-Cervalign, 1 Level, 10mm Material Number: 66-110… and 40 more products Locking mechanism has been observed to disassociate either intraoperatively or postoperatively resulting in replacement/revision |
| Sep 24, 2020 | Class II | Z-0299-2021 | Streamline MIS Navigation System Taps, Catalog Nos.
04-CANNDRILL-MDN
04-CANNTAP-40-MD… and 1 more product The Streamline Navigation System taps have experienced binding with other devices. |
| Sep 24, 2020 | Class II | Z-0322-2021 | Streamline OCT Navigated Instruments, Catalog numbers 26-DRILL-MDN
26-FTAP-30-MDN
26-FTA… System does not include a drill guide. |
| Sep 11, 2020 | Class II | Z-0320-2021 | Streamline MIS Navigation System Taps, Catalog numbers
04-CANNDRILL-MDN
04-CANNTAP-40… The Streamline MIS Navigation System taps are not compatible with the Streamline MIS tap sleeve. |
| Sep 16, 2016 | Class II | Z-0290-2017 | RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm… RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline¿ OCT Straight Rod, CoCr. 3.2mm (D) x 240mm… |
| Apr 22, 2016 | Class II | Z-1829-2016 | Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal… Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which… |
| Mar 18, 2016 | Class II | Z-1393-2016 | RTI Surgical Streamline Occipito-Cervico-Thoracic (OCT) Final Driver, Common Name: Screw… Multiple lots of the 26-FNLDRIVER, may have reduced mechanical strength, resulting in driver tip fracturing during use. |
| May 29, 2015 | Class II | Z-2053-2015 | Songer Spinal Cable System, Single Cable, Crimp, Bar, Stainless Steel, Sterile.
The… The box end label is incorrectly labeled as Titanium instead of Stainless Steel on one lot of implantable Songer Spinal Cable System, Single… |
| May 21, 2015 | Class II | Z-1916-2015 | Bullet-Tip PEEK VBR/IBF System, 32mm(l) 13mm(H) Implant 32-13-32 UDI (01)008464680341… Three lots of the Bullet -Tip VBR 13mm x 32 mm implants are mislabeled with a light green colored label, used for 16 mm trial instead of a… |
| Jun 11, 2014 | Class II | Z-2076-2014 | Tritium Sternal Plate System, Screw, 02.7mm X 12mm, 12 Pack, Sterile.
The Pioneer… Lack of Sterility Assurance |
FDA record score
8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
18 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jun 14, 2024 | 6 x 0.6 | 3.6 |
| Class II recall Z-1204-2021, this FEI as the recalling firm | Jan 15, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0299-2021, this FEI as the recalling firm | Sep 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0322-2021, this FEI as the recalling firm | Sep 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0320-2021, this FEI as the recalling firm | Sep 11, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0290-2017, this FEI as the recalling firm | Sep 16, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 23, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1829-2016, this FEI as the recalling firm | Apr 22, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1393-2016, this FEI as the recalling firm | Mar 18, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2053-2015, this FEI as the recalling firm | May 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1916-2015, this FEI as the recalling firm | May 21, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 11, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2076-2014, this FEI as the recalling firm | Jun 11, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0627-2014, this FEI as the recalling firm | Oct 7, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2616-2014, this FEI as the recalling firm | Jun 14, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2573-2014, this FEI as the recalling firm | Sep 10, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 3, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Apr 22, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Pioneer Surgical Technology, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Pioneer Surgical Technology, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Pioneer Surgical Technology, Inc. → Check this supplier
Facility details
- Company
- Pioneer Surgical Technology Inc.
- Location
- Marquette, MI (US)
- Source
- device