ENCISION, INC. FDA Inspection Record

FEI 1000117361 · Inspections, Form 483s, warning letters and import refusals

ENCISION, INC. is an FDA-registered facility (FEI 1000117361) located in BOULDER, CO, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 20, 2019: No Action Indicated (NAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
1, latest Jun 15, 2012

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 20, 2019No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Oct 19, 2015No Action Indicated (NAI)DHR content (21 CFR 820.184)
Lack of or inadequate process validation (21 CFR 820.75(a))
Dec 7, 2012Voluntary Action Indicated (VAI)Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Nonconforming product control (21 CFR 820.90(a))
Complaints (21 CFR 820.198(a))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
and 1 more
Dec 2, 2009Voluntary Action Indicated (VAI)Design plans- updated (21 CFR 820.30(b))
Design verification - Lack of or inadequate procedures (21 CFR 820.30(f))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Design validation - user needs and intended uses (21 CFR 820.30(g))
and 5 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jun 15, 2012Class IIZ-1951-2012AEM Disposable Electrodes, Device, Electrosurgical, Cutting and Coagulation and…
Encision is performing a field correction to their AEM Disposable Electrodes due to two consumer complaints reporting unanticipated alarms…

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

3 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIDec 7, 201210 years or older, so it adds no points0
Class II recall Z-1951-2012, this FEI as the recalling firmJun 15, 201210 years or older, so it adds no points0
Inspection classified VAIDec 2, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ENCISION, INC. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for ENCISION, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about ENCISION, INC. → Check this supplier

Facility details

Company
Encision Inc.
Location
BOULDER, CO (US)
Source
device