THE METRIX COMPANY FDA Inspection Record
THE METRIX COMPANY is an FDA-registered facility (FEI 1000117446) located in DUBUQUE, IA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Apr 25, 2019: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (2 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Oct 9, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 25, 2019 | Voluntary Action Indicated (VAI) | No citations listed |
| May 6, 2016 | Voluntary Action Indicated (VAI) | Design review - Lack of or inadequate procedures (21 CFR 820.30(e)) Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) Analyzing service report (21 CFR 820.200(b)) and 4 more |
| Jul 9, 2014 | No Action Indicated (NAI) | No citations listed |
| Apr 22, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Oct 9, 2024 | Class II | Z-0330-2025 | VYGON 100 mL EMPTY EVA CONTAINER With 2 Ports, REF CMS-9601; An empty single-use pouch… and 15 more products A limited number of IV bags have been found to leak during filling. |
| Sep 21, 2018 | Class II | Z-0001-2020 | SECURE EVA - DUAL CHAMBER CONTAINER With Screw Connectors, 1500mL, (legless) REF 66615… and 7 more products Potential for leaking product |
| Sep 13, 2017 | Class II | Z-3240-2017 | SECURE(R) 100mL EMPTY VINYL CONTAINER With 2 Ports - (5 Boxes of 40), REF 58719, STERILE… The hot stamp label on the bag incorrectly states the material of the vinyl (bag) container. The bags are correctly made with PVC material… |
| Jul 7, 2016 | Class II | V-0017-2017 | Metrix Bleeding Bag 15 G Needle, Part # 2000, non-sterile, not for human use, packed in… and 10 more products The hub of the needle on affected products may contain aluminum flakes. |
| Jun 23, 2016 | Class II | Z-2257-2016 | AUTOCOMP6 XPS High Speed Compounder REF 58810 and 7 more products An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when… |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0330-2025, this FEI as the recalling firm | Oct 9, 2024 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Apr 25, 2019 | 6 x 0.25 | 1.5 |
| Class II recall Z-0001-2020, this FEI as the recalling firm | Sep 21, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-3240-2017, this FEI as the recalling firm | Sep 13, 2017 | 4 x 0.25 | 1 |
| Class II recall V-0017-2017, this FEI as the recalling firm | Jul 7, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2257-2016, this FEI as the recalling firm | Jun 23, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 6, 2016 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 17, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |