THE METRIX COMPANY FDA Inspection Record

FEI 1000117446 · Inspections, Form 483s, warning letters and import refusals

THE METRIX COMPANY is an FDA-registered facility (FEI 1000117446) located in DUBUQUE, IA, US. RegDataLab tracks 4 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Apr 25, 2019: Voluntary Action Indicated (VAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Oct 9, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 25, 2019Voluntary Action Indicated (VAI)No citations listed
May 6, 2016Voluntary Action Indicated (VAI)Design review - Lack of or inadequate procedures (21 CFR 820.30(e))
Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
Analyzing service report (21 CFR 820.200(b))
and 4 more
Jul 9, 2014No Action Indicated (NAI)No citations listed
Apr 22, 2010No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Oct 9, 2024Class IIZ-0330-2025VYGON 100 mL EMPTY EVA CONTAINER With 2 Ports, REF CMS-9601; An empty single-use pouch… and 15 more products
A limited number of IV bags have been found to leak during filling.
Sep 21, 2018Class IIZ-0001-2020SECURE EVA - DUAL CHAMBER CONTAINER With Screw Connectors, 1500mL, (legless) REF 66615… and 7 more products
Potential for leaking product
Sep 13, 2017Class IIZ-3240-2017SECURE(R) 100mL EMPTY VINYL CONTAINER With 2 Ports - (5 Boxes of 40), REF 58719, STERILE…
The hot stamp label on the bag incorrectly states the material of the vinyl (bag) container. The bags are correctly made with PVC material…
Jul 7, 2016Class IIV-0017-2017Metrix Bleeding Bag 15 G Needle, Part # 2000, non-sterile, not for human use, packed in… and 10 more products
The hub of the needle on affected products may contain aluminum flakes.
Jun 23, 2016Class IIZ-2257-2016AUTOCOMP6 XPS High Speed Compounder REF 58810 and 7 more products
An issue was identified after a redesign of the auto compounding device. The compounder cannot meet the designed accuracy specification when…

FDA record score

6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 18, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0330-2025, this FEI as the recalling firmOct 9, 20244 x 0.62.4
Inspection classified VAIApr 25, 20196 x 0.251.5
Class II recall Z-0001-2020, this FEI as the recalling firmSep 21, 20184 x 0.251
Class II recall Z-3240-2017, this FEI as the recalling firmSep 13, 20174 x 0.251
Class II recall V-0017-2017, this FEI as the recalling firmJul 7, 201610 years or older, so it adds no points0
Class II recall Z-2257-2016, this FEI as the recalling firmJun 23, 201610 years or older, so it adds no points0
Inspection classified VAIMay 6, 201610 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 17, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 17, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 17, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 17, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 17, 2026Newest record Nov 16, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing THE METRIX COMPANY as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check THE METRIX COMPANY yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
The Metrix Company
Location
DUBUQUE, IA (US)
Source
device