Boston Scientific Corporation FDA Inspection Record
Boston Scientific Corporation is an FDA-inspected facility (FEI 1000120754) located in Lowell, MA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 9, 2014: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (3 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 1, latest Jun 2, 2015
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 9, 2014 | Voluntary Action Indicated (VAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Mar 18, 2011 | Voluntary Action Indicated (VAI) | Design validation - software validation documentation (21 CFR 820.30(g)) Design plans- updated (21 CFR 820.30(b)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) |
| Aug 21, 2009 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 2, 2015 | Class II | Z-1817-2015 | CLEARSIGN II Amplifier for LABSYSTEM PRO EP Recording System, 120 channels, using… Boston Scientific has received complaints indicating that, when using the blood pressure (BP) channels on the CLEARSIGN II Amplifier, the… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1817-2015, this FEI as the recalling firm | Jun 2, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 9, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 18, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 21, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Boston Scientific Corporation as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.