OAKWORKS, INC. FDA Inspection Record

FEI 1000121370 · Inspections, Form 483s, warning letters and import refusals

OAKWORKS, INC. is an FDA-registered facility (FEI 1000121370) located in NEW FREEDOM, PA, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jan 17, 2024: Voluntary Action Indicated (VAI)
Inspections on record
3 (3 VAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Dec 5, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jan 17, 2024Voluntary Action Indicated (VAI)Quality audits - auditor independence (21 CFR 820.22)
Lack of or inadequate procedures (21 CFR 820.100(a))
Complaints (21 CFR 820.198(a))
Design validation - simulated testing (21 CFR 820.30(g))
and 2 more
Mar 16, 2017Voluntary Action Indicated (VAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
Nonconforming product control (21 CFR 820.90(a))
DHR content (21 CFR 820.184)
and 6 more
May 2, 2014Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Design control - no procedures (21 CFR 820.30(a))
and 11 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 5, 2023Class IIZ-0727-2024Ultrasound EA (Ergonomic Advantage) table versions; G2 (GENERAL 2) -An Ultrasound… and 3 more products
Backrest platform stress fracture - the table is not operational in an inclined position
Apr 21, 2022Class IIZ-1238-2022PX200 Emergency Relief Bed; Model No. PXEXEB362280A6TTZZH4C19 and 1 more product
Product label was printed without the UDI number and warning label was printed with the incorrect weight capacity. The correct weight limit…
Jun 29, 2021Class IIZ-2349-2021Vascular Table without Fowler (USV1) and Ultrasound Vascular Table with Fowler…
Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting…
May 2, 2019Class IIZ-1411-2019OAKWORKS Inc.Lithotripsy/Urology Table: CFLU401, Catalog Number: 75216-T01 - Product…
Table assembly error with incorrect actuators for the lateral and longitudinal table movements
Jul 31, 2017Class IIZ-0703-2018Foot Controls impacted for the following devices: OAKWORKS¿ Medical CFPM 300 Imaging…
The foot control used on the OAKWORKS¿ Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can…

FDA record score

13 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

8 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJan 17, 20246 x 0.63.6
Class II recall Z-0727-2024, this FEI as the recalling firmDec 5, 20234 x 0.62.4
Class II recall Z-1238-2022, this FEI as the recalling firmApr 21, 20224 x 0.62.4
Inspection classified VAIMar 16, 20176 x 0.251.5
Class II recall Z-2349-2021, this FEI as the recalling firmJun 29, 20214 x 0.251
Class II recall Z-1411-2019, this FEI as the recalling firmMay 2, 20194 x 0.251
Class II recall Z-0703-2018, this FEI as the recalling firmJul 31, 20174 x 0.251
Inspection classified VAIMay 2, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing OAKWORKS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check OAKWORKS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Oakworks Inc
Location
NEW FREEDOM, PA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.