EXCELSIOR MEDICAL LLC FDA Inspection Record
EXCELSIOR MEDICAL LLC is an FDA-registered facility (FEI 1000122362) located in NEPTUNE, NJ, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 20, 2016: No Action Indicated (NAI)
- Inspections on record
- 5 (2 OAI, 1 VAI, 2 NAI)
- Last OAI inspection
- Sep 29, 2015
- Warning letters
- 1 listed by FDA for this FEI, latest Nov 7, 2014
- Recalls as the recalling firm
- 2, latest Apr 28, 2016
- Import refusals
- 10 or more
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 20, 2016 | No Action Indicated (NAI) | No citations listed |
| Sep 29, 2015 | Official Action Indicated (OAI) | Investigation of device failures (21 CFR 820.198(c)) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Aug 21, 2014 | Official Action Indicated (OAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Personnel requirements, Lack of or inadequate requirements (21 CFR 820.70(d)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 5 more |
| May 14, 2012 | No Action Indicated (NAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Supplier oversight (21 CFR 820.50(a)(2)) |
| Jan 8, 2010 | Voluntary Action Indicated (VAI) | Establish the design and development plan (21 CFR 820.30(b)) Lack of or inadequate procedures - Acceptance activities (21 CFR 820.80(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) Equipment suitability & capability (21 CFR 820.72(a)) and 4 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 28, 2016 | Class II | Z-1989-2016 | Excelsior Disposable Syringe Heparin Lock Flush, USP syringe (5 mL Fill in 10 mL syringe)… Certain lots of Heparin Lock Flush, USP Syringes (5 ml Fill in 10 ml Syringe) failed Anti-Factor IIa assay for potency during testing of… |
| Jun 5, 2014 | Class II | Z-2070-2014 | Excelsior Medical Sterile Triple Lead Tubing Set and Sterile Heavy Duty Triple Lead Tubing… There is a potential for a leak. A molding defect in the connector that joins the three leads to the main flow line may allow fluid to… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| 15 import refusals in 2024 | Dec 31, 2024 | 2 x 1 | 2 |
| Class II recall Z-1989-2016, this FEI as the recalling firm | Apr 28, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Sep 29, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Nov 7, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 21, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2070-2014, this FEI as the recalling firm | Jun 5, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 8, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 16, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 16, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 16, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing EXCELSIOR MEDICAL LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check EXCELSIOR MEDICAL LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Excelsior Medical LLC
- Location
- NEPTUNE, NJ (US)
- Source
- device