Ortho-Clinical Diagnostics, Inc. FDA Inspection Record

FEI 1000122536 · Inspections, Form 483s, warning letters and import refusals

Ortho-Clinical Diagnostics, Inc. is an FDA-registered facility (FEI 1000122536) located in ATHENS, OH, US. RegDataLab tracks 6 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Aug 3, 2021: No Action Indicated (NAI)
Inspections on record
6 (2 VAI, 4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
9, latest Feb 1, 2023

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 3, 2021No Action Indicated (NAI)No citations listed
Jan 29, 2018Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a))
Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2))
and 1 more
May 21, 2013No Action Indicated (NAI)No citations listed
Jan 17, 2013Voluntary Action Indicated (VAI)Not approved or obsolete document retrieval (21 CFR 820.40(a))
Mar 16, 2011No Action Indicated (NAI)No citations listed
Sep 24, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 1, 2023Class IIZ-1308-2023MicroVue C1-Inhibitor Plus EIA for measuring the amount of functional C1-lnhibitor protein…
Kits with affected microtiter plates have the potential to produce low OD values, causing the run to fail, per the Validation requirements…
Aug 13, 2021Class IIZ-2477-2021Lyra Direct SARS-CoV-2 Assay Emergency Use Authorization- Ref: M124. The Lyra Direct…
Revised Instructions for Use: Potential for specimens from patients with high SARS-CoV-2 viral loads to be falsely interpreted as negative…
Apr 27, 2021Class IZ-1928-2021Lyra SARS-CoV-2 Assay (M120) in vitro diagnostic test reagent
In rare situations patients infected with high viral load of SARS CoV 2 may have samples that generate Ct values less than or equal to (d) 5…
Apr 4, 2017Class IIZ-1496-2018MicroVue CIC-C1q EIA, Model A001. An enzyme immunoassay for the detection of circulating…
There is a potential for the slope of the standard curve to fail to meet the assay validation requirements for the assay.
Oct 26, 2016Class IIIZ-1520-2018Quidel MicroVue Intact PTH EIA, Model 8044. An enzyme immunoassay for the quantification…
There may be a loss of separation between Calibrators A and B which may impact the assay calibration at the lower end of the assay.
Jul 14, 2016Class IIZ-1797-2018MicroVue BAP EIA, Model 8012, provides a quantitative measure of bone-specific alkaline…
There is a possibility of low volume and/or leaking standard bottles.
Nov 3, 2015Class IIIZ-1497-2018MicroVue CIC-C1q EIA, Model A001. The MicroVue CIC-C1q EIA is for detection of circulating… and 2 more products
A component of the kit was found to contain bacterial contamination.
Oct 16, 2014Class IIZ-0417-2015MicroVue Bb Plus EIA is packaged as a kit which contains all reagent components to perform…
Samples are quantitating incorrectly, with approximately 50% higher concentration values than previously reported and historically…
Dec 31, 2012Class IIIZ-0792-2013Diagnostic Hybrids, Inc., RhMK, Model #:49-0600, 49-0600A, 49-0102A, and 49-T075A…
RhMK product fungal contamination.

FDA record score

10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

8 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-1928-2021, this FEI as the recalling firmApr 27, 202115 x 0.253.8
Class II recall Z-1308-2023, this FEI as the recalling firmFeb 1, 20234 x 0.62.4
Inspection classified VAIJan 29, 20186 x 0.251.5
Class II recall Z-2477-2021, this FEI as the recalling firmAug 13, 20214 x 0.251
Class II recall Z-1496-2018, this FEI as the recalling firmApr 4, 20174 x 0.251
Class II recall Z-1797-2018, this FEI as the recalling firmJul 14, 201610 years or older, so it adds no points0
Class II recall Z-0417-2015, this FEI as the recalling firmOct 16, 201410 years or older, so it adds no points0
Inspection classified VAIJan 17, 201310 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Ortho-Clinical Diagnostics, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Ortho-Clinical Diagnostics, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Quidel Corporation
Location
ATHENS, OH (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.