UNITED STATES ENDOSCOPY GROUP, INC. FDA Inspection Record
UNITED STATES ENDOSCOPY GROUP, INC. is an FDA-registered facility (FEI 1000122656) located in Mentor, OH, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Dec 7, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (3 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 10, latest Feb 4, 2019
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Dec 7, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) Documentation (21 CFR 820.100(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 2 more |
| Sep 14, 2012 | Voluntary Action Indicated (VAI) | No citations listed |
| Jul 21, 2011 | Voluntary Action Indicated (VAI) | Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3)) Lack of or inadequate procedures (21 CFR 820.100(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 4, 2019 | Class II | Z-1011-2019 | US Endoscopy Padlock Clip defect closure system
Ref: C910001
The Padlock Clip is… Potential esophageal laceration during a patient procedure |
| Dec 13, 2016 | Class III | Z-1416-2017 | Roth Net Polyp Retrieval The firm was notified that the expiration date on the outer carton is different than the expiration date on the inner package. The… |
| May 18, 2016 | Class II | Z-0289-2017 | Vari-Safe Injection Needle US endoscopy 5976 Heisley Road Mentor, Ohio 44060
A sterile… US Endoscopy is conducting a voluntary product recall of one lot of the Vari-Safe Injection Needle as it was identified that the lot was… |
| Jul 14, 2015 | Class II | Z-0001-2016 | Histolock Resection Device, US Endoscopy.
An electrosurgical device designed to be… The company has determined the sterility cannot be assured for the affected lot. |
| Mar 10, 2015 | Class II | Z-1453-2015 | Talon Grasping Device, 160 cm, US endoscopy. Used for retrieval of foreign bodies, tissue… A wire component on the distal grasping assembly of the device became detached.. |
| Feb 17, 2015 | Class II | Z-1943-2015 | Velocity" Biopsy Valve, 25 units per box,
Product Usage:
The disposable Velocity… The company has received reports of structural non conformance at the connection between the valve body and the irrigation tube, resulting… |
| Oct 31, 2014 | Class II | Z-0861-2015 | UroSeal Adjustable Endoscopic Valve. Model/Catalog # 00913410
The primary packaging is a… Packaging non-conformance related to the integrity of the sterile pouch seal. |
| Mar 7, 2013 | Class II | Z-1099-2013 | Torrent" irrigation tubing.
Product Usage: The Torrent irrigation system (tubing… The firm was notified by their customers that there was a water leakage from the white tubing connector. |
| Dec 22, 2011 | Class II | Z-2396-2012 | Infinity¿ sampling device, Cytology Brush, United States Endoscopy Group, Inc., 5976… US Endoscopy received five complaints which noted difficulty in deploying the cytology brush through the catheter. |
| Nov 14, 2011 | Class II | Z-0006-2013 | Endoscopic retrieval device.
The disposable Raptor grasping device in intended to… When the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is… |
FDA record score
3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
12 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Dec 7, 2016 | 6 x 0.25 | 1.5 |
| Class II recall Z-1011-2019, this FEI as the recalling firm | Feb 4, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0289-2017, this FEI as the recalling firm | May 18, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0001-2016, this FEI as the recalling firm | Jul 14, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1453-2015, this FEI as the recalling firm | Mar 10, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1943-2015, this FEI as the recalling firm | Feb 17, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0861-2015, this FEI as the recalling firm | Oct 31, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1099-2013, this FEI as the recalling firm | Mar 7, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 14, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2396-2012, this FEI as the recalling firm | Dec 22, 2011 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0006-2013, this FEI as the recalling firm | Nov 14, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 21, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 17, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing UNITED STATES ENDOSCOPY GROUP, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check UNITED STATES ENDOSCOPY GROUP, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- US Endoscopy Group Inc
- Location
- Mentor, OH (US)
- Source
- device