VOLK OPTICAL INC. FDA Inspection Record
VOLK OPTICAL INC. is an FDA-registered facility (FEI 1000122772) located in Mentor, OH, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 28, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 4 (2 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 4, latest Feb 2, 2016
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 28, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Documentation (21 CFR 820.80(c)) Supplier oversight (21 CFR 820.50(a)(2)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 1 more |
| Nov 6, 2014 | No Action Indicated (NAI) | No citations listed |
| Oct 24, 2012 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) |
| Jun 26, 2009 | No Action Indicated (NAI) | Change records, content (21 CFR 820.40(b)) Process control procedures (21 CFR 820.70(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 2, 2016 | Class III | Z-1177-2016 | Volk G-4 Small Ring, No Flange, Gonio Lens with case, UPC: 811608013602
for use as… The firm discovered that the incorrect lot number was engraved on the product. |
| Feb 27, 2015 | Class III | Z-1685-2015 | Volk 25D Large Clear BIO Lens with case UPC: 8 11608 01010 6, VOLK OPTICAL INC. The Volk… Mislabeling. The incorrect lot number was engraved on the anodized aluminum ring. |
| Feb 27, 2015 | Class III | Z-2055-2015 | Lombart 90D Clear with case
Visualization during diagnosis and laser therapy of the human… Mislabeling. The incorrect lot number was engraved on the anodized aluminum ring. The correct lot number is BE03356; however, BD03356 was… |
| Oct 25, 2012 | Class III | Z-0966-2013 | 20 Diopter Binocular Indirect Ophthalmoscope (BIO) Lens
For visualization during… The defect was with mis-engraving of the lot # directly on the product and its packaging. Mixed up device engraving of V20LC (lot #… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
2 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Sep 28, 2017 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Oct 24, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing VOLK OPTICAL INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check VOLK OPTICAL INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Volk Optical Inc
- Location
- Mentor, OH (US)
- Source