TRINITY STERILE, INC. FDA Inspection Record
TRINITY STERILE, INC. is an FDA-registered facility (FEI 1000123297) located in SALISBURY, MD, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- May 23, 2024: No Action Indicated (NAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Dec 13, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| May 23, 2024 | No Action Indicated (NAI) | No citations listed |
| Apr 26, 2019 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) |
| Mar 13, 2017 | Voluntary Action Indicated (VAI) | DHR - not or inadequately maintained (21 CFR 820.184) Design control - no procedures (21 CFR 820.30(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) and 5 more |
| Feb 13, 2013 | Voluntary Action Indicated (VAI) | Procedures for product review,disposition lack of/inadequate (21 CFR 820.90(b)(1)) |
| Sep 1, 2010 | Voluntary Action Indicated (VAI) | Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Procedures for product review,disposition lack of/inadequate (21 CFR 820.90(b)(1)) Quality Audit/Reaudit - conducted (21 CFR 820.22) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 13, 2024 | Class II | Z-1012-2025 | Brand Name: IMed Products
Product Name: Dressing Change Kit with ChloraPrep 3mL &… and 4 more products Potential for hole in package compromising sterility of the medical kit. |
| Jun 29, 2022 | Class II | Z-1548-2022 | Sklar Instruments IV Cath Dressing Tray Cs/25
Re-Order Number: 96-1708 and 3 more products Kits mislabeled as latex-free, the packaging for component 94-7002 Non-Adherent Dressing contains natural rubber latex. Potential risk to to… |
| Feb 23, 2021 | Class II | Z-1468-2021 | Midline Catheter Dressing Change Kit with Statlock - Product Usage: Intended to provide… Evidence of face masks or other material in the seal in several kits, causing improper seals and compromised sterility |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1012-2025, this FEI as the recalling firm | Dec 13, 2024 | 4 x 1 | 4 |
| Class II recall Z-1548-2022, this FEI as the recalling firm | Jun 29, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Apr 26, 2019 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Mar 13, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-1468-2021, this FEI as the recalling firm | Feb 23, 2021 | 4 x 0.25 | 1 |
| Inspection classified VAI | Feb 13, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 1, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 17, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 17, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 17, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 17, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 17, 2026 | Newest record Nov 16, 2026 | Current |
| Import refusals | Nov 15, 2026 | Newest record Nov 5, 2026 | Current |
Reviewing TRINITY STERILE, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check TRINITY STERILE, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Trinity Sterile, Inc.
- Location
- SALISBURY, MD (US)
- Source