Marina Medical Instruments, Inc. FDA Inspection Record

FEI 1000124817 ยท Inspections, Form 483s, warning letters and import refusals

Marina Medical Instruments, Inc. is an FDA-registered facility (FEI 1000124817) located in Davie, FL, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Apr 3, 2017: Voluntary Action Indicated (VAI)
Inspections on record
8 (2 OAI, 4 VAI, 2 NAI)
Last OAI inspection
Sep 1, 2011
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Apr 21, 2009. Many letters do not state an FEI: search warning letters by company name
Import refusals
1

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Apr 3, 2017Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
Jan 9, 2014Voluntary Action Indicated (VAI)Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Jan 9, 2014Voluntary Action Indicated (VAI)Software validation for automated processes (21 CFR 820.70(i))
Report of Death or Serious Injury (21 CFR 803.50(a)(1))
Jan 9, 2014No Action Indicated (NAI)Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Periodic equipment inspections (21 CFR 820.70(g)(2))
Lack of or inadequate final acceptance procedures (21 CFR 820.80(d))
Lack of or inadequate process validation (21 CFR 820.75(a))
May 2, 2012Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Quality System Procedures (21 CFR 820.20(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
and 2 more
May 2, 2012No Action Indicated (NAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Documentation (21 CFR 820.80(e))
Nonconforming product control (21 CFR 820.90(a))
and 5 more
Sep 1, 2011Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate procedures (21 CFR 820.100(a))
Document change records, maintained. (21 CFR 820.40(b))
Supplier notification of changes (21 CFR 820.50(b))
and 4 more
Dec 5, 2008Official Action Indicated (OAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design control - no procedures (21 CFR 820.30(a))
Design history file (21 CFR 820.30(j))
Documentation (21 CFR 820.75(a))
and 10 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 27, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIApr 3, 20176 x 0.251.5
Inspection classified VAIJan 9, 201410 years or older, so it adds no points0
Inspection classified VAIJan 9, 201410 years or older, so it adds no points0
Inspection classified VAIMay 2, 201210 years or older, so it adds no points0
Inspection classified OAISep 1, 201110 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsApr 21, 200910 years or older, so it adds no points0
Inspection classified OAIDec 5, 200810 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Marina Medical Instruments, Inc. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Marina Medical Instruments, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Marina Medical Instruments, Inc. โ†’ Check this supplier

Facility details

FDA Legal Name
MARINA MEDICAL INSTRUMENTS, INC.
Company
Marina Medical Instruments Inc.
Location
Davie, FL (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.