VENTANA MEDICAL SYSTEMS, INC. FDA Inspection Record

FEI 1000125183 · Inspections, Form 483s, warning letters and import refusals

VENTANA MEDICAL SYSTEMS, INC. is an FDA-registered facility (FEI 1000125183) located in Tucson, AZ, US. RegDataLab tracks 12 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 23, 2022: No Action Indicated (NAI)
Inspections on record
12 (12 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Form 483 and inspection documents
1 in FDA's reading room
Recalls as the recalling firm
10, latest Apr 20, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Showing the 10 most recent of 12 inspections. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 23, 2022No Action Indicated (NAI)No citations listed
Sep 15, 2022No Action Indicated (NAI)No citations listed
Jul 29, 2021No Action Indicated (NAI)No citations listed
Feb 26, 2020No Action Indicated (NAI)No citations listed
Aug 14, 2019No Action Indicated (NAI)No citations listed
Jul 13, 2017No Action Indicated (NAI)No citations listed
Jul 14, 2016No Action Indicated (NAI)No citations listed
Mar 27, 2015No Action Indicated (NAI)No citations listed
Jul 17, 2012No Action Indicated (NAI)No citations listed
Jan 19, 2012No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Apr 20, 2026Class IIZ-2913-2026navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital…
Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window
Jun 27, 2024Class IIZ-2644-2024Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false…
Nov 18, 2022Class IIZ-0907-2023VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD)…
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test…
Jun 1, 2022Class IIZ-1374-2022Ventana HE 600 System, automated slide preparer, for use in laboratories.
There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
Jan 20, 2022Class IIZ-0894-2022BenchMark ULTRA and DISCOVERY ULTRA Instruments
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's…
Oct 4, 2019Class IIZ-2262-2020Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended…
The firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for…
Aug 2, 2018Class IZ-2996-2018OptiView DAB IHC Detection Kit, Catalog Number 06396500001, model 760-700… and 20 more products
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
Dec 19, 2017Class IZ-0729-2018OptiView DAB IHC Detection Kit, Ventana Part Number 760-700, Roche GMMI 06396500001 in… and 11 more products
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and…
Sep 28, 2012Class IIZ-0833-2013VENTANA (a member of the Roche Group) BenchMark ULTRA; REF/MU 750-600; Research Use… and 1 more product
Potiential for a waste fluid overflow condition to accur while running decontamination cycles on the BenchMark ULTRA automated staining…
Jun 12, 2008Class IIZ-2118-2013Benchmark XT/LT and Discovery XT Model Numbers: N750-BMK-FS, N750-BMKXT-FS…
Ventana Medical System is initiating the recall of the BenchMark and DISCOVERY series instruments because there is a possibility that the…

Form 483s and inspection documents

  • 483 on FDA.gov
    FDA listing date Feb 10, 2026, which is not always the inspection or issue date. Published by FDA Oct 22, 2026.

FDA record score

22 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2913-2026, this FEI as the recalling firmApr 20, 20264 x 14
Class I recall Z-2996-2018, this FEI as the recalling firmAug 2, 201815 x 0.253.8
Class I recall Z-0729-2018, this FEI as the recalling firmDec 19, 201715 x 0.253.8
Class II recall Z-2644-2024, this FEI as the recalling firmJun 27, 20244 x 0.62.4
Class II recall Z-0907-2023, this FEI as the recalling firmNov 18, 20224 x 0.62.4
Class II recall Z-1374-2022, this FEI as the recalling firmJun 1, 20224 x 0.62.4
Class II recall Z-0894-2022, this FEI as the recalling firmJan 20, 20224 x 0.62.4
Class II recall Z-2262-2020, this FEI as the recalling firmOct 4, 20194 x 0.251
Class II recall Z-0833-2013, this FEI as the recalling firmSep 28, 201210 years or older, so it adds no points0
Class II recall Z-2118-2013, this FEI as the recalling firmJun 12, 200810 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing VENTANA MEDICAL SYSTEMS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check VENTANA MEDICAL SYSTEMS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Ventana Medical Systems, Inc.
Location
Tucson, AZ (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.