ZIMMER MANUFACTURING B.V. FDA Inspection Record

FEI 1000132037 · Inspections, Form 483s, warning letters and import refusals

ZIMMER MANUFACTURING B.V. is an FDA-registered facility (FEI 1000132037) located in MERCEDITA, PR, US. RegDataLab tracks 8 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Oct 1, 2020: No Action Indicated (NAI)
Inspections on record
8 (4 OAI, 2 VAI, 2 NAI)
Last OAI inspection
Jun 11, 2014
Warning letters
1 listed by FDA for this FEI, latest Sep 19, 2012
Recalls as the recalling firm
8, latest May 4, 2016

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 1, 2020No Action Indicated (NAI)No citations listed
Jul 20, 2017Voluntary Action Indicated (VAI)Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b))
Documentation (21 CFR 820.100(b))
Lack of or inadequate process validation (21 CFR 820.75(a))
Nov 17, 2015Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Lack of or inadequate process validation (21 CFR 820.75(a))
Jun 11, 2014Official Action Indicated (OAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Procedures not adequately established or maintained (21 CFR 820.40)
Lack/Inad procedure-Monitoring/Control of Validated Proces (21 CFR 820.75(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 5 more
Aug 16, 2013Official Action Indicated (OAI)No citations listed
May 17, 2013Official Action Indicated (OAI)No citations listed
Jun 22, 2012Official Action Indicated (OAI)Documentation (21 CFR 820.100(b))
Lack of or inadequate process validation (21 CFR 820.75(a))
Feb 27, 2009No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 4, 2016Class IIZ-1673-2016Zimmer Plates and Screws (ZPS) 3.5 mm Cortical Screw SelfTapping, 10 mm Length… and 11 more products
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Mar 17, 2016Class IIZ-1340-2016Persona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left… and 1 more product
The firm received a customer complaint establishing that the product (Persona PS Articular Surfaces)information shown on the box differed by…
Jan 11, 2016Class IIZ-1099-2016Product 1 consists of all product under , product code: JWH and same usage: Item… and 71 more products
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The…
Dec 3, 2015Class IIZ-0590-2016NexGEN System REF 5980-37-01 Tibial Component Size Stemmed 3 Precoat, and REF 5986-37-01…
A product complaint was received stating that a carton labeled as a Precoat Stemmed Tibia Size 3 was opened for use and the contents were a…
Dec 2, 2015Class IIZ-0449-2016Label reading in part as: "NexGen System, Complete Knee Solution; Reference Number…
Complaints were received from Thailand reporting that there was no white paper (Tyvek) seal on the inner package of P/N:00-5980-037-01 unit…
Dec 4, 2014Class IIZ-0875-2015Trilogy¿ Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is…
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor…
Nov 21, 2014Class IIZ-0824-2015Bone Screw, self-tapping The Trilogy shells are available in three configurations; uni…
Analysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one…
Jun 12, 2014Class IIZ-2032-2014Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD…
Two lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number…

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

15 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIJul 20, 20176 x 0.251.5
Class II recall Z-1673-2016, this FEI as the recalling firmMay 4, 201610 years or older, so it adds no points0
Class II recall Z-1340-2016, this FEI as the recalling firmMar 17, 201610 years or older, so it adds no points0
Class II recall Z-1099-2016, this FEI as the recalling firmJan 11, 201610 years or older, so it adds no points0
Class II recall Z-0590-2016, this FEI as the recalling firmDec 3, 201510 years or older, so it adds no points0
Class II recall Z-0449-2016, this FEI as the recalling firmDec 2, 201510 years or older, so it adds no points0
Inspection classified VAINov 17, 201510 years or older, so it adds no points0
Class II recall Z-0875-2015, this FEI as the recalling firmDec 4, 201410 years or older, so it adds no points0
Class II recall Z-0824-2015, this FEI as the recalling firmNov 21, 201410 years or older, so it adds no points0
Class II recall Z-2032-2014, this FEI as the recalling firmJun 12, 201410 years or older, so it adds no points0
Inspection classified OAIJun 11, 201410 years or older, so it adds no points0
Inspection classified OAIAug 16, 201310 years or older, so it adds no points0
Inspection classified OAIMay 17, 201310 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsSep 19, 201210 years or older, so it adds no points0
Inspection classified OAIJun 22, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ZIMMER MANUFACTURING B.V. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ZIMMER MANUFACTURING B.V. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Zimmer Manufacturing B.V.
Location
MERCEDITA, PR (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.