HAAG-STREIT USA, INC. FDA Inspection Record
HAAG-STREIT USA, INC. is an FDA-registered facility (FEI 1000136533) located in MASON, OH, US. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Mar 7, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 6, latest Nov 3, 2021
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 7, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Records of complaint investigation (21 CFR 820.198(e)) Lack of System for Determining MDR Events (21 CFR 803.17(a)(2)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 3, 2021 | Class II | Z-0477-2022 | OCT-Camera 211 01 A1, Running software iOCT-Control 3 v1.4 and 1 more product Malfunction of the automatic laser beam shut-off, the OCT unit might not recognize whether the laser beam is safely switched off |
| Apr 15, 2021 | Class II | Z-1560-2021 | Eyesuite 9.3.1 software, LENSTAR, MODEL LS900: ac-powered slit-lamp biomicroscope. Examination data and patient name may be mixed up when printing or generating a pdf with the interface at Pacific Coast and Laser Institute… |
| Nov 20, 2020 | Class II | Z-0697-2021 | Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515)… Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it… |
| Jul 31, 2019 | Class II | Z-2482-2019 | HS Hi-R NEO 900
Reference # 657 820
Serial # from 101
Product Usage:
A wide range… and 2 more products software error in the central control unit of the floor stand. |
| Jan 5, 2017 | Class II | Z-1500-2017 | EyeSuite i.8.2.1.0 Software for ophthalmic use including selection of Intra Ocular Lenses… There is a possibility for data to be stored under the wrong patient on the DICOM Server following a non-standard workflow. |
| Aug 17, 2012 | Class II | Z-2411-2012 | Octopus 101 Perimeter; PeriTrend Software Analysis
A perimeter is a device intended to… The firm recalled the device after learning of a possible deviation in the stimulus luminance of the Perimeter OCTOPUS 101. |
FDA record score
7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 7, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0477-2022, this FEI as the recalling firm | Nov 3, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1560-2021, this FEI as the recalling firm | Apr 15, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0697-2021, this FEI as the recalling firm | Nov 20, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2482-2019, this FEI as the recalling firm | Jul 31, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1500-2017, this FEI as the recalling firm | Jan 5, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2411-2012, this FEI as the recalling firm | Aug 17, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing HAAG-STREIT USA, INC. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for HAAG-STREIT USA, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about HAAG-STREIT USA, INC. → Check this supplier
Facility details
- Company
- Haag-Streit USA Inc
- Location
- MASON, OH (US)
- Source