Microline Surgical, Inc FDA Inspection Record
Microline Surgical, Inc is an FDA-registered facility (FEI 1000137656) located in BEVERLY, MA, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Sep 18, 2013: No Action Indicated (NAI)
- Inspections on record
- 4 (1 OAI, 2 VAI, 1 NAI)
- Last OAI inspection
- Feb 25, 2010
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 10, 2010. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Nov 5, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 18, 2013 | No Action Indicated (NAI) | No citations listed |
| Jul 21, 2011 | Voluntary Action Indicated (VAI) | Mixups (21 CFR 820.120(d)) |
| Feb 8, 2011 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) |
| Feb 25, 2010 | Official Action Indicated (OAI) | Procedures (21 CFR 820.100(b)) Quality Audits - defined intervals (21 CFR 820.22) General (21 CFR 820.198(a)) Report of risk to health (21 CFR 806.10(a)(1)) and 1 more |
Warning letters
- Aug 10, 2010: Warning letter listed on FDA's Data Dashboard. View compliance record (113148).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 5, 2024 | Class II | Z-0663-2025 | Microscissor Tip (Disposable); Microline Catalog #: 3122. A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off. |
| Mar 22, 2019 | Class II | Z-1698-2019 | TLS2 Thermal Ligating Shears, Catalog Number 132-131D The shears exhibited a high rate of complete or intermittent failure to actuate. As a result, they do not perform their intended function… |
| Nov 1, 2017 | Class II | Z-1740-2018 | ReNew Dolphin Nose Disposable Dissector Tip, Catalog Number: 3312, UDI… and 1 more product Certain lots of disposable laparoscopic instrument tips may potentially have a compromised sterile barrier. |
| Jul 13, 2017 | Class II | Z-0507-2018 | The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic… The heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off… |
| May 12, 2014 | Class II | Z-1986-2014 | Microline Surgical ReNew Laproscopic Fenestrated Grasper Forceps Tips
Ref 3221 and 1 more product Potential for the Fenestrated Grasper Tip jaw to break when force is applied to the jaw |
FDA record score
5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0663-2025, this FEI as the recalling firm | Nov 5, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1698-2019, this FEI as the recalling firm | Mar 22, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1740-2018, this FEI as the recalling firm | Nov 1, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0507-2018, this FEI as the recalling firm | Jul 13, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1986-2014, this FEI as the recalling firm | May 12, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 21, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 8, 2011 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Aug 10, 2010 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Feb 25, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Microline Surgical, Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Microline Surgical, Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Microline Surgical, Inc.
- Location
- BEVERLY, MA (US)
- Source
- device