Microline Surgical, Inc FDA Inspection Record

FEI 1000137656 · Inspections, Form 483s, warning letters and import refusals

Microline Surgical, Inc is an FDA-registered facility (FEI 1000137656) located in BEVERLY, MA, US. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Sep 18, 2013: No Action Indicated (NAI)
Inspections on record
4 (1 OAI, 2 VAI, 1 NAI)
Last OAI inspection
Feb 25, 2010
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Aug 10, 2010. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
5, latest Nov 5, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 18, 2013No Action Indicated (NAI)No citations listed
Jul 21, 2011Voluntary Action Indicated (VAI)Mixups (21 CFR 820.120(d))
Feb 8, 2011Voluntary Action Indicated (VAI)Documentation (21 CFR 820.100(b))
Feb 25, 2010Official Action Indicated (OAI)Procedures (21 CFR 820.100(b))
Quality Audits - defined intervals (21 CFR 820.22)
General (21 CFR 820.198(a))
Report of risk to health (21 CFR 806.10(a)(1))
and 1 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Nov 5, 2024Class IIZ-0663-2025Microscissor Tip (Disposable); Microline Catalog #: 3122.
A lot of ReNew Microscissor Tips exhibited a 2% rate of failure in terms of the heat shrink breaking off.
Mar 22, 2019Class IIZ-1698-2019TLS2 Thermal Ligating Shears, Catalog Number 132-131D
The shears exhibited a high rate of complete or intermittent failure to actuate. As a result, they do not perform their intended function…
Nov 1, 2017Class IIZ-1740-2018ReNew Dolphin Nose Disposable Dissector Tip, Catalog Number: 3312, UDI… and 1 more product
Certain lots of disposable laparoscopic instrument tips may potentially have a compromised sterile barrier.
Jul 13, 2017Class IIZ-0507-2018The ReNew Reusable Grasper and Dissector Tips are to be used with the ReNew Laparoscopic…
The heat-shrink insulation tube may split during autoclave sterilization/re-sterilization. If a split occurs, the heat-shrink could fall off…
May 12, 2014Class IIZ-1986-2014Microline Surgical ReNew Laproscopic Fenestrated Grasper Forceps Tips Ref 3221 and 1 more product
Potential for the Fenestrated Grasper Tip jaw to break when force is applied to the jaw

FDA record score

5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 26, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0663-2025, this FEI as the recalling firmNov 5, 20244 x 0.62.4
Class II recall Z-1698-2019, this FEI as the recalling firmMar 22, 20194 x 0.251
Class II recall Z-1740-2018, this FEI as the recalling firmNov 1, 20174 x 0.251
Class II recall Z-0507-2018, this FEI as the recalling firmJul 13, 20174 x 0.251
Class II recall Z-1986-2014, this FEI as the recalling firmMay 12, 201410 years or older, so it adds no points0
Inspection classified VAIJul 21, 201110 years or older, so it adds no points0
Inspection classified VAIFeb 8, 201110 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsAug 10, 201010 years or older, so it adds no points0
Inspection classified OAIFeb 25, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 26, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 26, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 26, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 26, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Microline Surgical, Inc as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Microline Surgical, Inc yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Microline Surgical, Inc.
Location
BEVERLY, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.