St. Jude Medical, Atrial Fibrillation Division, Inc. FDA Inspection Record

FEI 1000139754 · Inspections, Form 483s, warning letters and import refusals

St. Jude Medical, Atrial Fibrillation Division, Inc. is an FDA-registered facility (FEI 1000139754) located in Saint Paul, MN, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Mar 29, 2019: No Action Indicated (NAI)
Inspections on record
3 (3 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
4, latest Dec 18, 2023
Import refusals
1

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 29, 2019No Action Indicated (NAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design plans - Lack of or inadequate (21 CFR 820.30(b))
Design output - documentation (21 CFR 820.30(d))
May 17, 2018No Action Indicated (NAI)No citations listed
Sep 28, 2017No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 18, 2023Class IIZ-0814-2024TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an…
When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product…
May 5, 2021Class IIZ-1853-2021v.1.2 Upgrade Kit or v.1.2 upgrade installed on WorkMate Claris System, Model H700123… and 2 more products
Due to a software error, the user may lose functionality of the system or the screen may turn black during operation.
Apr 22, 2021Class IIZ-1549-2021NeuroTherm NT2000iX Radiofrequency (RF) Generator, Model Number RFG-NT-2000
Potential capacitor failure which typically manifests as in inability to provide output on one or more channels. In some instances, users…
Apr 5, 2019Class IIZ-1493-2019TactiSys Quartz Equipment (with software version 1.7) Model: PN-004 400 Part Numbers…
In reported cases, the device log on the TactiSys Quartz Equipment operating on Software Version 1.7.0 fills the allocated disk space, which…

FDA record score

5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

4 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0814-2024, this FEI as the recalling firmDec 18, 20234 x 0.62.4
Class II recall Z-1853-2021, this FEI as the recalling firmMay 5, 20214 x 0.251
Class II recall Z-1549-2021, this FEI as the recalling firmApr 22, 20214 x 0.251
Class II recall Z-1493-2019, this FEI as the recalling firmApr 5, 20194 x 0.251

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing St. Jude Medical, Atrial Fibrillation Division, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check St. Jude Medical, Atrial Fibrillation Division, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
St. Jude Medical, Atrial Fibrillation Division, Inc.
Location
Saint Paul, MN (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.