Terumo Cardiovascular Systems Corp FDA Inspection Record
Terumo Cardiovascular Systems Corp is an FDA-inspected facility (FEI 1000149028) located in Ashland, MA, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 25, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Feb 3, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 25, 2017 | Voluntary Action Indicated (VAI) | Complaints (21 CFR 820.198(a)) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) |
| Jan 17, 2013 | Voluntary Action Indicated (VAI) | DHR - not or inadequately maintained (21 CFR 820.184) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate procedures for labeling (21 CFR 820.120) |
| May 7, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 3, 2022 | Class II | Z-0949-2022 | Cardiovascular Procedure Kit catalog # 76645 & 73806 Potential for Cardiovascular Procedure Kit packaging damage that occurred during shipping. |
| Dec 18, 2017 | Class II | Z-1314-2018 | Terumo Cardiovascular Procedure Kits containing Pall LG6NS LeukoGuard¿ Leukocyte Reduction… Possible blood leaks through the hydrophobic portion of the Pall LG6NS LeukoGuard Leukocyte Reduction Arterial Blood Filters. |
| May 12, 2017 | Class II | Z-2550-2018 | Tubing Pack Convenience Kit, Product Code 65217
Product Usage:
Intended for use only… Kit was labeled with the incorrect Expiration Date of April 30, 2019. |
| Jul 25, 2016 | Class II | Z-2706-2016 | Terumo Custom Cardiovascular System Cardiovascular Procedure Kits labeled containing… Terumo Custom Cardiovascular Procedure Kits are labeled as Xcoating surface coated tubing, but tubing is non-coated |
| Aug 24, 2015 | Class II | Z-2786-2015 | Terumo Custom Cardiovascular Procedure Kit Tubing Pack
P/N: 66041-1
Terumo… and 6 more products Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators |
| May 6, 2015 | Class II | Z-1828-2015 | Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves.
For use only in the… The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is… |
| Mar 28, 2014 | Class II | Z-1465-2014 | Terumo Cardiovascular Procedure Kit- X-Coated Pump and Table Pack with FX25RE
Catalog… Incorrect customer notification label applied to the Tyvek lid of this specific lot of Terumo¿ Cardiovascular Procedure Kits |
| Mar 1, 2011 | Class II | Z-0156-2015 | Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass
Item… and 5 more products Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the… |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0949-2022, this FEI as the recalling firm | Feb 3, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Sep 25, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-1314-2018, this FEI as the recalling firm | Dec 18, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2550-2018, this FEI as the recalling firm | May 12, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2706-2016, this FEI as the recalling firm | Jul 25, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2786-2015, this FEI as the recalling firm | Aug 24, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1828-2015, this FEI as the recalling firm | May 6, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1465-2014, this FEI as the recalling firm | Mar 28, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 17, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0156-2015, this FEI as the recalling firm | Mar 1, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Terumo Cardiovascular Systems Corp as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Terumo Cardiovascular Systems Corp yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Terumo Cardiovascular Systems Corp
- Location
- Ashland, MA (US)
- Source
- inspections