Terumo Cardiovascular Systems Corp FDA Inspection Record

FEI 1000149028 · Inspections, Form 483s, warning letters and import refusals

Terumo Cardiovascular Systems Corp is an FDA-inspected facility (FEI 1000149028) located in Ashland, MA, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 25, 2017: Voluntary Action Indicated (VAI)
Inspections on record
3 (2 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
8, latest Feb 3, 2022

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 25, 2017Voluntary Action Indicated (VAI)Complaints (21 CFR 820.198(a))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
Jan 17, 2013Voluntary Action Indicated (VAI)DHR - not or inadequately maintained (21 CFR 820.184)
Lack of or inadequate procedures (21 CFR 820.100(a))
Lack of or inadequate procedures for labeling (21 CFR 820.120)
May 7, 2010No Action Indicated (NAI)No citations listed

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 3, 2022Class IIZ-0949-2022Cardiovascular Procedure Kit catalog # 76645 & 73806
Potential for Cardiovascular Procedure Kit packaging damage that occurred during shipping.
Dec 18, 2017Class IIZ-1314-2018Terumo Cardiovascular Procedure Kits containing Pall LG6NS LeukoGuard¿ Leukocyte Reduction…
Possible blood leaks through the hydrophobic portion of the Pall LG6NS LeukoGuard Leukocyte Reduction Arterial Blood Filters.
May 12, 2017Class IIZ-2550-2018Tubing Pack Convenience Kit, Product Code 65217 Product Usage: Intended for use only…
Kit was labeled with the incorrect Expiration Date of April 30, 2019.
Jul 25, 2016Class IIZ-2706-2016Terumo Custom Cardiovascular System Cardiovascular Procedure Kits labeled containing…
Terumo Custom Cardiovascular Procedure Kits are labeled as Xcoating surface coated tubing, but tubing is non-coated
Aug 24, 2015Class IIZ-2786-2015Terumo Custom Cardiovascular Procedure Kit Tubing Pack P/N: 66041-1 Terumo… and 6 more products
Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators
May 6, 2015Class IIZ-1828-2015Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the…
The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is…
Mar 28, 2014Class IIZ-1465-2014Terumo Cardiovascular Procedure Kit- X-Coated Pump and Table Pack with FX25RE Catalog…
Incorrect customer notification label applied to the Tyvek lid of this specific lot of Terumo¿ Cardiovascular Procedure Kits
Mar 1, 2011Class IIZ-0156-2015Terumo Custom Cardioplegia Delivery Set - Heat Exchanger Cardiopulmonary Bypass Item… and 5 more products
Defect in the luer connector can compromise the seal on the female side of the connector and the compromised seal can introduce air in the…

FDA record score

6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-0949-2022, this FEI as the recalling firmFeb 3, 20224 x 0.62.4
Inspection classified VAISep 25, 20176 x 0.251.5
Class II recall Z-1314-2018, this FEI as the recalling firmDec 18, 20174 x 0.251
Class II recall Z-2550-2018, this FEI as the recalling firmMay 12, 20174 x 0.251
Class II recall Z-2706-2016, this FEI as the recalling firmJul 25, 201610 years or older, so it adds no points0
Class II recall Z-2786-2015, this FEI as the recalling firmAug 24, 201510 years or older, so it adds no points0
Class II recall Z-1828-2015, this FEI as the recalling firmMay 6, 201510 years or older, so it adds no points0
Class II recall Z-1465-2014, this FEI as the recalling firmMar 28, 201410 years or older, so it adds no points0
Inspection classified VAIJan 17, 201310 years or older, so it adds no points0
Class II recall Z-0156-2015, this FEI as the recalling firmMar 1, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Terumo Cardiovascular Systems Corp as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Terumo Cardiovascular Systems Corp yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Terumo Cardiovascular Systems Corp
Location
Ashland, MA (US)
Source
inspections