NRT X-RAY A/S FDA Inspection Record
NRT X-RAY A/S is an FDA-registered facility (FEI 1000188474) located in HASSELAGER Midtjylland, DK. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 24, 2023: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Apr 17, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 24, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Documentation (21 CFR 820.100(b)) Certification label or tag (21 CFR 1010.2(b)) |
| Jul 31, 2015 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) Failure to report (21 CFR 1002.20(a)) FDA not notified of defect or noncompliance (21 CFR 1003.10(a)) |
| Jan 22, 2010 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 17, 2024 | Class II | Z-1784-2024 | Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop… X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is… |
| Apr 27, 2023 | Class II | Z-1668-2023 | Intelli-C, Right side suspended Tabletop, REF: 03400000 and 1 more product There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector… |
| Jan 21, 2022 | Class II | Z-0526-2022 | 1. Footrest ( use exclusively in combination with Intelli-C devices)
2. Shoulder support… Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a… |
FDA record score
11 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
5 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 24, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-1784-2024, this FEI as the recalling firm | Apr 17, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-1668-2023, this FEI as the recalling firm | Apr 27, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0526-2022, this FEI as the recalling firm | Jan 21, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Jan 22, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing NRT X-RAY A/S as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check NRT X-RAY A/S yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- NRT X-RAY A/S
- Location
- HASSELAGER Midtjylland (DK)
- Source