CLINICAL INNOVATIONS, LLC FDA Inspection Record
CLINICAL INNOVATIONS, LLC is an FDA-registered facility (FEI 1000220650) located in MURRAY, UT, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 15, 2019: Voluntary Action Indicated (VAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Mar 9, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 15, 2019 | Voluntary Action Indicated (VAI) | Lack of or inadequate process validation (21 CFR 820.75(a)) |
| May 8, 2013 | No Action Indicated (NAI) | No citations listed |
| Apr 27, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 9, 2026 | Class II | Z-1945-2026 | Brand Name: Kiwi Omni Vacuum Delivery System
Product Name: Kiwi Omni Vacuum Delivery… Due to complaints of device breakage at the traction force gauge to handle joint. |
| Jul 19, 2022 | Class II | F-1682-2022 | Sweetums 24% Sucrose Solution Preservative Free 15 mL cups VPSU15037288. Each box… Firm received complaints of leaking cups, black liquid, and black particulates in the 24% Sucrose Solution. |
| Jul 19, 2022 | Class II | F-1684-2022 | Sweetums 24% sucrose solution preservative free. Vials containing 1ml or 2ml of 24%… and 1 more product Firm received complaints of leaking, black liquid, and black particulates in the product vials. |
| Aug 13, 2020 | Class II | Z-0541-2021 | Clinical Innovations ROM Plus Collection Kit
IFU:
ROM Plus Rupture of Membrane Test… Possible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into… |
| Aug 6, 2019 | Class II | Z-2072-2020 | Kiwi Complete Vacuum Delivery System, Model No. VAC-6000MTE - Product Usage: The Kiwi is a… There is a potential for loss of vacuum or failure to generate vacuum when using the firm's fetal vacuum delivery system. |
| May 30, 2019 | Class II | Z-2118-2020 | Clearview Total Uterine Manipulator (model UM750), The device is labeled in part… An incorrect component used during manufacturing. The uterine manipulator tip support does not have the "notch" required to attach the… |
| Jan 5, 2015 | Class II | Z-0123-2016 | The Ebb Complete Tamponade System, model #CTS-1000 and Code #OQY.
Intended for use in… Clinical Innovations, LLC is voluntarily recalling the Ebb Complete Tamponade System (CTS-1000) Lot Number 1214-F-403.) to address the… |
| Dec 30, 2014 | Class II | Z-1014-2015 | Ebb Complete Tamponade System, BD-OTS Obstetric Catheter, REF CTS-1000, Rx Only, Sterile… Potential for a balloon leak in all lots of (CTS-1000) manufactured since August 2014. |
FDA record score
13 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1945-2026, this FEI as the recalling firm | Mar 9, 2026 | 4 x 1 | 4 |
| Class II recall F-1682-2022, this FEI as the recalling firm | Jul 19, 2022 | 4 x 0.6 | 2.4 |
| Class II recall F-1684-2022, this FEI as the recalling firm | Jul 19, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 15, 2019 | 6 x 0.25 | 1.5 |
| Class II recall Z-0541-2021, this FEI as the recalling firm | Aug 13, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2072-2020, this FEI as the recalling firm | Aug 6, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2118-2020, this FEI as the recalling firm | May 30, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0123-2016, this FEI as the recalling firm | Jan 5, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1014-2015, this FEI as the recalling firm | Dec 30, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing CLINICAL INNOVATIONS, LLC as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check CLINICAL INNOVATIONS, LLC yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Clinical Innovations, LLC
- Location
- MURRAY, UT (US)
- Source
- device