ORTHO-CLINICAL DIAGNOSTICS, INC. FDA Inspection Record

FEI 1000305840 · Inspections, Form 483s, warning letters and import refusals

ORTHO-CLINICAL DIAGNOSTICS, INC. is an FDA-registered facility (FEI 1000305840) located in ROCHESTER, NY, US. RegDataLab tracks 1 FDA inspection at this facility.

What FDA found at this site

Latest inspection
Mar 27, 2017: Voluntary Action Indicated (VAI)
Inspections on record
1 (1 VAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
19, latest Dec 19, 2024

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Mar 27, 2017Voluntary Action Indicated (VAI)Design validation - Risk analysis (21 CFR 820.30(g))
Procedures not adequately established or maintained (21 CFR 820.40)
Lack of or inadequate procedures for handling (21 CFR 820.140)
Complaints (21 CFR 820.198(a))
and 1 more

Recalls

Showing the 10 most recent of 19 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 19, 2024Class IIZ-1005-2025Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number… and 2 more products
The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board…
Oct 25, 2023Class IIZ-0510-2024VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry…
Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included…
Sep 12, 2023Class IIZ-0236-2024VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per…
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results…
Aug 5, 2023Class IIZ-2623-2023VITROS XT Chemistry Products TBIL-ALKP Slides and 1 more product
Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS…
May 3, 2021Class IIZ-1688-2021VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent…
Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when…
Mar 26, 2021Class IIZ-1489-2021VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference…
Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the…
Jul 13, 2020Class IIZ-2873-2020VITROS Chemistry Products TBIL Slides- 5 PACK/300 SLDS In vitro Diagnostic quantitatively… and 2 more products
Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides
May 11, 2020Class IIZ-2364-2020VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/300 SLDS, Product Code… and 3 more products
Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within…
Apr 1, 2020Class IIZ-1930-2020VITROS Chemistry Products CKMB Slides, Creatine Kinase MB, Product Code 805 8232-300… and 1 more product
The Spanish and Swedish language versions of the IFU state that the assay can be tested with Plasma which is false and could result in…
Sep 10, 2019Class IIZ-0267-2020VITROS chemistry Products Cl- Slides, Catalog No. 6844471 (which supports Urine CL-)… and 2 more products
Users may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the…

FDA record score

24 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 24, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

19 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1005-2025, this FEI as the recalling firmDec 19, 20244 x 14
Class II recall Z-0510-2024, this FEI as the recalling firmOct 25, 20234 x 0.62.4
Class II recall Z-0236-2024, this FEI as the recalling firmSep 12, 20234 x 0.62.4
Class II recall Z-2623-2023, this FEI as the recalling firmAug 5, 20234 x 0.62.4
Inspection classified VAIMar 27, 20176 x 0.251.5
Class II recall Z-1688-2021, this FEI as the recalling firmMay 3, 20214 x 0.251
Class II recall Z-1489-2021, this FEI as the recalling firmMar 26, 20214 x 0.251
Class II recall Z-2873-2020, this FEI as the recalling firmJul 13, 20204 x 0.251
Class II recall Z-2364-2020, this FEI as the recalling firmMay 11, 20204 x 0.251
Class II recall Z-1930-2020, this FEI as the recalling firmApr 1, 20204 x 0.251
Class II recall Z-0267-2020, this FEI as the recalling firmSep 10, 20194 x 0.251
Class II recall Z-2202-2019, this FEI as the recalling firmJun 17, 20194 x 0.251
Class II recall Z-3280-2018, this FEI as the recalling firmMay 5, 20184 x 0.251
Class II recall Z-0273-2018, this FEI as the recalling firmJun 12, 20174 x 0.251
Class II recall Z-1117-2017, this FEI as the recalling firmDec 20, 20164 x 0.251
Class II recall Z-1179-2017, this FEI as the recalling firmDec 8, 20164 x 0.251
Class II recall Z-2273-2016, this FEI as the recalling firmMay 18, 201610 years or older, so it adds no points0
Class II recall Z-0848-2015, this FEI as the recalling firmNov 6, 201410 years or older, so it adds no points0
Class II recall Z-0007-2015, this FEI as the recalling firmAug 20, 201410 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 24, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 24, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 24, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 24, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing ORTHO-CLINICAL DIAGNOSTICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check ORTHO-CLINICAL DIAGNOSTICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Ortho-Clinical Diagnostics, INc.
Location
ROCHESTER, NY (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.