ORTHO-CLINICAL DIAGNOSTICS, INC. FDA Inspection Record
ORTHO-CLINICAL DIAGNOSTICS, INC. is an FDA-registered facility (FEI 1000305841) located in Rochester, NY, US. RegDataLab tracks 1 FDA inspection at this facility.
What FDA found at this site
- Latest inspection
- Mar 31, 2017: Voluntary Action Indicated (VAI)
- Inspections on record
- 1 (1 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 11, latest Feb 27, 2024
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 31, 2017 | Voluntary Action Indicated (VAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 27, 2024 | Class II | Z-1372-2024 | VITROS Performance Verifier I and 1 more product The issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to… |
| May 8, 2023 | Class II | Z-2129-2023 | VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only.
VITROS… Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG… |
| Dec 11, 2018 | Class II | Z-0848-2019 | VITROS Chemistry Products VALP Reagent Pack. For in vitro diagnostic use only.
CATALOG… Negative bias results when using VITROS VALP Reagent, GEN 25. |
| Sep 22, 2017 | Class II | Z-0911-2018 | VITROS Chemistry Products DAT Performance Verifier I
in vitro diagnostic and 4 more products This product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower… |
| Jan 5, 2017 | Class III | Z-1181-2017 | VITROS Chemistry Products FS Calibrator 1, REF 680 1873 12 x 3 mL, Rx ONLY, IVD; used in… The company received a customer complaint for the inability to calibrate VITROS Chemistry Products dLDL Reagent when using VITROS FS… |
| May 24, 2016 | Class III | Z-2119-2016 | VITROS Chemistry Products Calibrator Kit 29, REF/Product Code 680 2344, used in… Ortho Clinical Diagnostics started receiving customer complaints for biased results when using VITROS dTIBC Reagent GEN 30 product. Ortho's… |
| Apr 19, 2016 | Class III | Z-2804-2016 | The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with… There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the… |
| Sep 28, 2015 | Class III | Z-0202-2016 | VITROS Chemistry Products Calibrator Kit 9 Calibrator, 2 mL, Product Code 8568040, Unique… The firm is conducting a product correction due to an increase in U90-382 or 6LU condition codes (wash errors) when using VITROS Calibrator… |
| Jul 15, 2014 | Class II | Z-2214-2014 | VITROS¿ Chemistry Products CDM PROM Calibration Data Module Rev. No. 0168 and 0169… Inability to calibrate or use VITROS NH3 DT Slides, GEN 59 with VITROS Chemistry Products DT Calibrator Kit, Lot 44 when CDM PROM 0168 or… |
| May 12, 2014 | Class II | Z-1804-2014 | VITROS (R) Chemistry Products Calibrator Kit 17, Product Code/ Catalog Number 6801701… and 1 more product Calibrator values assigned for the calibrator kit lots were released using a reference method that was undergoing validation for a new lot… |
FDA record score
8 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1372-2024, this FEI as the recalling firm | Feb 27, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-2129-2023, this FEI as the recalling firm | May 8, 2023 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Mar 31, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0848-2019, this FEI as the recalling firm | Dec 11, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0911-2018, this FEI as the recalling firm | Sep 22, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2214-2014, this FEI as the recalling firm | Jul 15, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1804-2014, this FEI as the recalling firm | May 12, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1490-2014, this FEI as the recalling firm | Jan 29, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ORTHO-CLINICAL DIAGNOSTICS, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ORTHO-CLINICAL DIAGNOSTICS, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Ortho-Clinical Diagnostics
- Location
- Rochester, NY (US)
- Source
- device