New Star Lasers, Inc. FDA Inspection Record
New Star Lasers, Inc. is an FDA-inspected facility (FEI 1000307100) located in Roseville, CA, US. RegDataLab tracks 4 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 20, 2017: No Action Indicated (NAI)
- Inspections on record
- 4 (1 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Sep 19, 2016
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 20, 2017 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Documentation (21 CFR 820.100(b)) |
| Aug 24, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Jun 21, 2016 | No Action Indicated (NAI) | No citations listed |
| Nov 16, 2011 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) DHR content (21 CFR 820.184) Lack of procedures, or not maintained (21 CFR 820.40) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Sep 19, 2016 | Class II | Z-2914-2016 | CT3Plus (a.k.a. CT3PZ);
Catalog number: 7300-0038/ 7300-0038-1.
Dermatology: For use… and 2 more products FDA inspection found that customers who had received devices with promotional material with outside use indications were not notified when… |
| May 13, 2015 | Class II | Z-1918-2015 | StoneLight 30 Laser System (NS3000).
For use in surgical procedures using open… During internal testing a situation was discovered where when changing the settings, a new setting could not be accepted and potentially… |
FDA record score
0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
3 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2914-2016, this FEI as the recalling firm | Sep 19, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 24, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1918-2015, this FEI as the recalling firm | May 13, 2015 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing New Star Lasers, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check New Star Lasers, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- New Star Lasers, Inc.
- Location
- Roseville, CA (US)
- Source