SIEMENS MEDICAL SOLUTIONS USA, INC. FDA Inspection Record

FEI 1000514697 · Inspections, Form 483s, warning letters and import refusals

SIEMENS MEDICAL SOLUTIONS USA, INC. is an FDA-registered facility (FEI 1000514697) located in Hoffman Estates, IL, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jun 10, 2019: No Action Indicated (NAI)
Inspections on record
3 (1 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
27, latest Jul 12, 2023
Import refusals
1

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 10, 2019No Action Indicated (NAI)No citations listed
Sep 9, 2015No Action Indicated (NAI)No citations listed
Apr 27, 2009Voluntary Action Indicated (VAI)Time to report - 10 days (21 CFR 806.10(b))

Recalls

Showing the 10 most recent of 27 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 12, 2023Class IIZ-2256-2023The following Pro.specta Emission Computed Tomography Systems that were manufactured with…
During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger…
Jan 16, 2023Class IIZ-1134-2023SIEMENS Biograph Horizon 3R Mobile; System, Tomography, Computed, Emission and 2 more products
This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position.
Mar 11, 2022Class IIZ-0893-2022Symbia Intevo Bold and Symbia Intevo 6 SPECT/CT Systems
There is a quality issue with the three bolts that hold the CT balancing weights, which could create a safety issue.
Aug 14, 2020Class IIZ-2934-2020Symbia S-Series and Symbia T-Series - Product Usage: Intended use of the device SPECT…
This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in…
Nov 30, 2018Class IIZ-0707-2019Siemens Biograph Horizon PET/CT System, Product Usage: The Siemens Biograph systems…
Improper design or specifications.
Oct 5, 2018Class IIZ-0372-2019EVO Excel, Emission Computed Tomographic System, Material Number 10910501
The Evo Excel External-ECG Power-Connection current limit is mis-labeled as 1.0A. It should be labeled 0.55A.,
Sep 28, 2018Class IIZ-0177-2019Biograph Horizon a) 3R, Material Number 10532746 b) 4R, Material Number 10532748 The… and 1 more product
The terminals on the plug may become loose and overheat.
May 29, 2018Class IIZ-2248-2018Biograph mCT Systems: (a) Biograph mCT S(64)-3R, Material Number 10248669 Product…
Siemens Medical Solutions USA, Inc., Molecular Imaging business line became aware, through our supplier that some laser cover windows may…
Feb 21, 2018Class IIZ-1930-2018Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide…
Possibility that the Biograph Horizon systems performing CT retrospective cardiac gating or PET cardiac gating examinations may experience a…
Jan 26, 2018Class IIZ-1597-2018Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic…
Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is…

FDA record score

16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 22, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

28 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2256-2023, this FEI as the recalling firmJul 12, 20234 x 0.62.4
Class II recall Z-1134-2023, this FEI as the recalling firmJan 16, 20234 x 0.62.4
Class II recall Z-0893-2022, this FEI as the recalling firmMar 11, 20224 x 0.62.4
Class II recall Z-2934-2020, this FEI as the recalling firmAug 14, 20204 x 0.251
Class II recall Z-0707-2019, this FEI as the recalling firmNov 30, 20184 x 0.251
Class II recall Z-0372-2019, this FEI as the recalling firmOct 5, 20184 x 0.251
Class II recall Z-0177-2019, this FEI as the recalling firmSep 28, 20184 x 0.251
Class II recall Z-2248-2018, this FEI as the recalling firmMay 29, 20184 x 0.251
Class II recall Z-1930-2018, this FEI as the recalling firmFeb 21, 20184 x 0.251
Class II recall Z-1597-2018, this FEI as the recalling firmJan 26, 20184 x 0.251
Class II recall Z-0093-2018, this FEI as the recalling firmNov 7, 20174 x 0.251
Class II recall Z-0047-2018, this FEI as the recalling firmSep 18, 20174 x 0.251
Class II recall Z-2758-2016, this FEI as the recalling firmAug 5, 201610 years or older, so it adds no points0
Class II recall Z-2898-2020, this FEI as the recalling firmMar 31, 201610 years or older, so it adds no points0
Class II recall Z-1262-2016, this FEI as the recalling firmFeb 26, 201610 years or older, so it adds no points0
Class II recall Z-0687-2016, this FEI as the recalling firmDec 31, 201510 years or older, so it adds no points0
Class II recall Z-2435-2015, this FEI as the recalling firmJul 31, 201510 years or older, so it adds no points0
Class II recall Z-0208-2016, this FEI as the recalling firmJun 26, 201510 years or older, so it adds no points0
Class II recall Z-0911-2015, this FEI as the recalling firmAug 12, 201410 years or older, so it adds no points0
Class II recall Z-2712-2014, this FEI as the recalling firmJul 28, 201410 years or older, so it adds no points0
Class II recall Z-2042-2014, this FEI as the recalling firmApr 23, 201410 years or older, so it adds no points0
Class II recall Z-0914-2015, this FEI as the recalling firmApr 8, 201410 years or older, so it adds no points0
Class II recall Z-0960-2015, this FEI as the recalling firmMar 5, 201410 years or older, so it adds no points0
Class II recall Z-0115-2015, this FEI as the recalling firmOct 28, 201310 years or older, so it adds no points0
Class II recall Z-0061-2017, this FEI as the recalling firmAug 16, 201310 years or older, so it adds no points0
Class II recall Z-1086-2015, this FEI as the recalling firmOct 23, 201210 years or older, so it adds no points0
Class II recall Z-1296-2015, this FEI as the recalling firmApr 11, 201210 years or older, so it adds no points0
Inspection classified VAIApr 27, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 22, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 22, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 22, 2026Newest record Nov 13, 2026Current
Warning letter textNov 22, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 22, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing SIEMENS MEDICAL SOLUTIONS USA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check SIEMENS MEDICAL SOLUTIONS USA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Siemens Medical Solutions USA, Inc.
Location
Hoffman Estates, IL (US)
Source
device