SIEMENS MEDICAL SOLUTIONS USA, INC. FDA Inspection Record
SIEMENS MEDICAL SOLUTIONS USA, INC. is an FDA-registered facility (FEI 1000514697) located in Hoffman Estates, IL, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jun 10, 2019: No Action Indicated (NAI)
- Inspections on record
- 3 (1 VAI, 2 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 27, latest Jul 12, 2023
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 10, 2019 | No Action Indicated (NAI) | No citations listed |
| Sep 9, 2015 | No Action Indicated (NAI) | No citations listed |
| Apr 27, 2009 | Voluntary Action Indicated (VAI) | Time to report - 10 days (21 CFR 806.10(b)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 12, 2023 | Class II | Z-2256-2023 | The following Pro.specta Emission Computed Tomography Systems that were manufactured with… During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger… |
| Jan 16, 2023 | Class II | Z-1134-2023 | SIEMENS Biograph Horizon 3R Mobile; System, Tomography, Computed, Emission and 2 more products This is a potential problem with the locking mechanism that is used to secure the PET gantry in either its scanning or service position. |
| Mar 11, 2022 | Class II | Z-0893-2022 | Symbia Intevo Bold and Symbia Intevo 6 SPECT/CT Systems There is a quality issue with the three bolts that hold the CT balancing weights, which could create a safety issue. |
| Aug 14, 2020 | Class II | Z-2934-2020 | Symbia S-Series and Symbia T-Series - Product Usage:
Intended use of the device SPECT… This letter is to inform you of the possibility that your Symbia S or Symbia T system(s) may have unintended slow motion of the detectors in… |
| Nov 30, 2018 | Class II | Z-0707-2019 | Siemens Biograph Horizon PET/CT System,
Product Usage:
The Siemens Biograph systems… Improper design or specifications. |
| Oct 5, 2018 | Class II | Z-0372-2019 | EVO Excel, Emission Computed Tomographic System, Material Number 10910501 The Evo Excel External-ECG Power-Connection current limit is mis-labeled as 1.0A. It should be labeled 0.55A., |
| Sep 28, 2018 | Class II | Z-0177-2019 | Biograph Horizon
a) 3R, Material Number 10532746
b) 4R, Material Number 10532748
The… and 1 more product The terminals on the plug may become loose and overheat. |
| May 29, 2018 | Class II | Z-2248-2018 | Biograph mCT Systems:
(a) Biograph mCT S(64)-3R, Material Number 10248669
Product… Siemens Medical Solutions USA, Inc., Molecular Imaging business line became aware, through our supplier that some laser cover windows may… |
| Feb 21, 2018 | Class II | Z-1930-2018 | Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide… Possibility that the Biograph Horizon systems performing CT retrospective cardiac gating or PET cardiac gating examinations may experience a… |
| Jan 26, 2018 | Class II | Z-1597-2018 | Biograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic… Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is… |
FDA record score
16 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
28 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2256-2023, this FEI as the recalling firm | Jul 12, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1134-2023, this FEI as the recalling firm | Jan 16, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0893-2022, this FEI as the recalling firm | Mar 11, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-2934-2020, this FEI as the recalling firm | Aug 14, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0707-2019, this FEI as the recalling firm | Nov 30, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0372-2019, this FEI as the recalling firm | Oct 5, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0177-2019, this FEI as the recalling firm | Sep 28, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2248-2018, this FEI as the recalling firm | May 29, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1930-2018, this FEI as the recalling firm | Feb 21, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1597-2018, this FEI as the recalling firm | Jan 26, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0093-2018, this FEI as the recalling firm | Nov 7, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0047-2018, this FEI as the recalling firm | Sep 18, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2758-2016, this FEI as the recalling firm | Aug 5, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2898-2020, this FEI as the recalling firm | Mar 31, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1262-2016, this FEI as the recalling firm | Feb 26, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0687-2016, this FEI as the recalling firm | Dec 31, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2435-2015, this FEI as the recalling firm | Jul 31, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0208-2016, this FEI as the recalling firm | Jun 26, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0911-2015, this FEI as the recalling firm | Aug 12, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2712-2014, this FEI as the recalling firm | Jul 28, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2042-2014, this FEI as the recalling firm | Apr 23, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0914-2015, this FEI as the recalling firm | Apr 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0960-2015, this FEI as the recalling firm | Mar 5, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0115-2015, this FEI as the recalling firm | Oct 28, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0061-2017, this FEI as the recalling firm | Aug 16, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1086-2015, this FEI as the recalling firm | Oct 23, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1296-2015, this FEI as the recalling firm | Apr 11, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 27, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing SIEMENS MEDICAL SOLUTIONS USA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check SIEMENS MEDICAL SOLUTIONS USA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Siemens Medical Solutions USA, Inc.
- Location
- Hoffman Estates, IL (US)
- Source
- device