STAT MEDICAL DEVICES, INC. FDA Inspection Record
STAT MEDICAL DEVICES, INC. is an FDA-registered facility (FEI 1000525253) located in North Miami Beach, FL, US. RegDataLab tracks 7 FDA inspections and 2 compliance actions at this facility.
What FDA found at this site
- Latest inspection
- Mar 18, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 7 (2 OAI, 3 VAI, 2 NAI)
- Last OAI inspection
- Aug 21, 2013
- Warning letters
- 2 listed by FDA for this FEI, latest Jul 23, 2015
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 18, 2025 | Voluntary Action Indicated (VAI) | Label to bear a unique device identifier (21 CFR 801.20(a)) Design control - no procedures (21 CFR 820.30(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Training records (21 CFR 820.25(b)) |
| Nov 7, 2022 | No Action Indicated (NAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| May 20, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Feb 20, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Documentation (21 CFR 820.100(b)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Aug 21, 2013 | Official Action Indicated (OAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate organizational structure (21 CFR 820.20(b)) Evaluation of suppliers, contractors, etc., requirements (21 CFR 820.50(a)) and 1 more |
| Feb 18, 2010 | No Action Indicated (NAI) | No citations listed |
| Dec 5, 2008 | Official Action Indicated (OAI) | Design verification procedures (21 CFR 820.30(f)) Documentation (21 CFR 820.75(a)) Results of investigation (21 CFR 820.198(e)(6)) Design verification - documentation (21 CFR 820.30(f)) and 4 more |
FDA record score
6 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 18, 2025 | 6 x 1 | 6 |
| Inspection classified VAI | May 20, 2016 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Jul 23, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 20, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 21, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Apr 17, 2009 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Dec 5, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 14, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 2, 2026 | Newest record Oct 22, 2026 | Current |
Reviewing STAT MEDICAL DEVICES, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check STAT MEDICAL DEVICES, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Stat Medical Devices, Inc.
- Location
- North Miami Beach, FL (US)
- Source