C.R. BARD, INC. FDA Inspection Record
C.R. BARD, INC. is an FDA-registered facility (FEI 1018233) located in COVINGTON, GA, US. Registered operations: STERILIZE. RegDataLab tracks 8 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Nov 12, 2024: No Action Indicated (NAI)
- Inspections on record
- 8 (2 VAI, 6 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 56, latest Jun 25, 2026
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 12, 2024 | No Action Indicated (NAI) | No citations listed |
| Jan 28, 2020 | No Action Indicated (NAI) | No citations listed |
| Jun 8, 2018 | No Action Indicated (NAI) | No citations listed |
| Apr 19, 2016 | Voluntary Action Indicated (VAI) | Design review - documentation (21 CFR 820.30(e)) Investigation of device failures (21 CFR 820.198(c)) Documentation of validated process performance (21 CFR 820.75(b)(2)) Documentation (21 CFR 820.100(b)) and 1 more |
| Feb 5, 2014 | No Action Indicated (NAI) | No citations listed |
| Jul 25, 2012 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Apr 21, 2010 | No Action Indicated (NAI) | No citations listed |
| Mar 12, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 25, 2026 | Class II | Z-3114-2026 | Products that contain the Pericare Wipes:
Bardex Lubricath Foley Catheter… and 5 more products Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and… |
| Apr 23, 2026 | Class II | Z-2226-2026 | Silastic Foley Catheters
Silastic Brand Foley Catheter… Due to stain present on the surface of affected foley catheters. |
| Nov 6, 2025 | Class II | Z-0905-2026 | Bard InLay Optima Size: 6 Fr. x 14 cm, Ureteral Stent Kit
REF 787614 and 3 more products Labeling discrepancy; Ureteral stent actual size may not match the actual product label. |
| Jul 17, 2025 | Class II | Z-2589-2025 | SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter… and 2 more products Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with… |
| Jun 18, 2025 | Class II | Z-2462-2025 | Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone… Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or… |
| Jun 3, 2025 | Class II | Z-2074-2025 | BD Heyman Follower Straight Tip for following sizes:
10Fr - REF:21110,
12Fr… and 2 more products Cather packaging may contain the incorrect French size. |
| Apr 17, 2025 | Class I | Z-1755-2025 | Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, REF:0092220 18 Fr. (6mm, 7/32")… and 4 more products Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to… |
| May 9, 2024 | Class II | Z-2154-2024 | Bard Medical SureStep Foley Tray System, 16 Fr., Catalog #A319416AM; and Bard Medical… Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature… |
| Dec 20, 2023 | Class II | Z-0790-2024 | Bard Nasogastric Sump Tube with ENFit, 10Fr., 36in Long, REF EN0042100 and 16 more products BD (C.R. Bard) has received user complaints reporting that during use of BD Bard NG Sump Tubes with ENFit Connectors, inadequate or reduced… |
| Nov 15, 2023 | Class II | Z-0554-2024 | Sensica Urine Output System, Catalog Number SCCS1002 BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica… |
FDA record score
74 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
53 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-1755-2025, this FEI as the recalling firm | Apr 17, 2025 | 15 x 1 | 15 |
| Class II recall Z-3114-2026, this FEI as the recalling firm | Jun 25, 2026 | 4 x 1 | 4 |
| Class II recall Z-2226-2026, this FEI as the recalling firm | Apr 23, 2026 | 4 x 1 | 4 |
| Class II recall Z-0905-2026, this FEI as the recalling firm | Nov 6, 2025 | 4 x 1 | 4 |
| Class II recall Z-2589-2025, this FEI as the recalling firm | Jul 17, 2025 | 4 x 1 | 4 |
| Class II recall Z-2462-2025, this FEI as the recalling firm | Jun 18, 2025 | 4 x 1 | 4 |
| Class II recall Z-2074-2025, this FEI as the recalling firm | Jun 3, 2025 | 4 x 1 | 4 |
| Class II recall Z-2154-2024, this FEI as the recalling firm | May 9, 2024 | 4 x 0.6 | 2.4 |
| Class II recall Z-0790-2024, this FEI as the recalling firm | Dec 20, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0554-2024, this FEI as the recalling firm | Nov 15, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1698-2023, this FEI as the recalling firm | Apr 20, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1293-2022, this FEI as the recalling firm | May 19, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-0666-2022, this FEI as the recalling firm | Jan 27, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-2521-2021, this FEI as the recalling firm | Aug 5, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1431-2021, this FEI as the recalling firm | Jan 11, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1970-2020, this FEI as the recalling firm | Apr 15, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1219-2020, this FEI as the recalling firm | Jan 13, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0803-2019, this FEI as the recalling firm | Dec 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0800-2019, this FEI as the recalling firm | Dec 6, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2536-2018, this FEI as the recalling firm | Jun 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2943-2018, this FEI as the recalling firm | Mar 21, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1354-2018, this FEI as the recalling firm | Feb 27, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2064-2018, this FEI as the recalling firm | Dec 13, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0565-2018, this FEI as the recalling firm | Nov 10, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2243-2018, this FEI as the recalling firm | Nov 10, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1241-2019, this FEI as the recalling firm | Nov 8, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0271-2018, this FEI as the recalling firm | Oct 16, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0488-2018, this FEI as the recalling firm | Sep 29, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1307-2018, this FEI as the recalling firm | Sep 14, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-3226-2017, this FEI as the recalling firm | Jul 28, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2872-2017, this FEI as the recalling firm | Jun 2, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-3042-2017, this FEI as the recalling firm | Mar 31, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1871-2017, this FEI as the recalling firm | Mar 9, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0267-2018, this FEI as the recalling firm | Dec 9, 2016 | 4 x 0.25 | 1 |
| Class II recall Z-0403-2017, this FEI as the recalling firm | Oct 13, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0363-2017, this FEI as the recalling firm | Oct 12, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0585-2017, this FEI as the recalling firm | Oct 12, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0600-2017, this FEI as the recalling firm | Oct 5, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0735-2017, this FEI as the recalling firm | Aug 3, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0967-2017, this FEI as the recalling firm | Jul 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2830-2016, this FEI as the recalling firm | Jul 11, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 19, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1395-2016, this FEI as the recalling firm | Mar 4, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1357-2016, this FEI as the recalling firm | Dec 18, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0760-2016, this FEI as the recalling firm | Nov 10, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1617-2016, this FEI as the recalling firm | Nov 9, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2732-2015, this FEI as the recalling firm | Aug 24, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2384-2015, this FEI as the recalling firm | Jul 7, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1358-2015, this FEI as the recalling firm | Jan 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0163-2015, this FEI as the recalling firm | Oct 7, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0010-2013, this FEI as the recalling firm | Sep 7, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 25, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2327-2012, this FEI as the recalling firm | Jun 29, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing C.R. BARD, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check C.R. BARD, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- C.R. Bard Inc
- Location
- COVINGTON, GA (US)
- Registration Expires
- Dec 31, 2027
- Operations
- STERILIZE
- Source
- both