C.R. BARD, INC. FDA Inspection Record

FEI 1018233 · Inspections, Form 483s, warning letters and import refusals

C.R. BARD, INC. is an FDA-registered facility (FEI 1018233) located in COVINGTON, GA, US. Registered operations: STERILIZE. RegDataLab tracks 8 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Nov 12, 2024: No Action Indicated (NAI)
Inspections on record
8 (2 VAI, 6 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
56, latest Jun 25, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Nov 12, 2024No Action Indicated (NAI)No citations listed
Jan 28, 2020No Action Indicated (NAI)No citations listed
Jun 8, 2018No Action Indicated (NAI)No citations listed
Apr 19, 2016Voluntary Action Indicated (VAI)Design review - documentation (21 CFR 820.30(e))
Investigation of device failures (21 CFR 820.198(c))
Documentation of validated process performance (21 CFR 820.75(b)(2))
Documentation (21 CFR 820.100(b))
and 1 more
Feb 5, 2014No Action Indicated (NAI)No citations listed
Jul 25, 2012Voluntary Action Indicated (VAI)Documentation (21 CFR 820.100(b))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Apr 21, 2010No Action Indicated (NAI)No citations listed
Mar 12, 2010No Action Indicated (NAI)No citations listed

Recalls

Showing the 10 most recent of 56 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jun 25, 2026Class IIZ-3114-2026Products that contain the Pericare Wipes: Bardex Lubricath Foley Catheter… and 5 more products
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and…
Apr 23, 2026Class IIZ-2226-2026Silastic Foley Catheters Silastic Brand Foley Catheter…
Due to stain present on the surface of affected foley catheters.
Nov 6, 2025Class IIZ-0905-2026Bard InLay Optima Size: 6 Fr. x 14 cm, Ureteral Stent Kit REF 787614 and 3 more products
Labeling discrepancy; Ureteral stent actual size may not match the actual product label.
Jul 17, 2025Class IIZ-2589-2025SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter… and 2 more products
Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with…
Jun 18, 2025Class IIZ-2462-2025Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone…
Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or…
Jun 3, 2025Class IIZ-2074-2025BD Heyman Follower Straight Tip for following sizes: 10Fr - REF:21110, 12Fr… and 2 more products
Cather packaging may contain the incorrect French size.
Apr 17, 2025Class IZ-1755-2025Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, REF:0092220 18 Fr. (6mm, 7/32")… and 4 more products
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to…
May 9, 2024Class IIZ-2154-2024Bard Medical SureStep Foley Tray System, 16 Fr., Catalog #A319416AM; and Bard Medical…
Some of the devices may function incorrectly resulting in no temperature display or erratic temperatures being displayed on the temperature…
Dec 20, 2023Class IIZ-0790-2024Bard Nasogastric Sump Tube with ENFit, 10Fr., 36in Long, REF EN0042100 and 16 more products
BD (C.R. Bard) has received user complaints reporting that during use of BD Bard NG Sump Tubes with ENFit Connectors, inadequate or reduced…
Nov 15, 2023Class IIZ-0554-2024Sensica Urine Output System, Catalog Number SCCS1002
BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica…

FDA record score

74 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

53 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-1755-2025, this FEI as the recalling firmApr 17, 202515 x 115
Class II recall Z-3114-2026, this FEI as the recalling firmJun 25, 20264 x 14
Class II recall Z-2226-2026, this FEI as the recalling firmApr 23, 20264 x 14
Class II recall Z-0905-2026, this FEI as the recalling firmNov 6, 20254 x 14
Class II recall Z-2589-2025, this FEI as the recalling firmJul 17, 20254 x 14
Class II recall Z-2462-2025, this FEI as the recalling firmJun 18, 20254 x 14
Class II recall Z-2074-2025, this FEI as the recalling firmJun 3, 20254 x 14
Class II recall Z-2154-2024, this FEI as the recalling firmMay 9, 20244 x 0.62.4
Class II recall Z-0790-2024, this FEI as the recalling firmDec 20, 20234 x 0.62.4
Class II recall Z-0554-2024, this FEI as the recalling firmNov 15, 20234 x 0.62.4
Class II recall Z-1698-2023, this FEI as the recalling firmApr 20, 20234 x 0.62.4
Class II recall Z-1293-2022, this FEI as the recalling firmMay 19, 20224 x 0.62.4
Class II recall Z-0666-2022, this FEI as the recalling firmJan 27, 20224 x 0.62.4
Class II recall Z-2521-2021, this FEI as the recalling firmAug 5, 20214 x 0.251
Class II recall Z-1431-2021, this FEI as the recalling firmJan 11, 20214 x 0.251
Class II recall Z-1970-2020, this FEI as the recalling firmApr 15, 20204 x 0.251
Class II recall Z-1219-2020, this FEI as the recalling firmJan 13, 20204 x 0.251
Class II recall Z-0803-2019, this FEI as the recalling firmDec 13, 20184 x 0.251
Class II recall Z-0800-2019, this FEI as the recalling firmDec 6, 20184 x 0.251
Class II recall Z-2536-2018, this FEI as the recalling firmJun 13, 20184 x 0.251
Class II recall Z-2943-2018, this FEI as the recalling firmMar 21, 20184 x 0.251
Class II recall Z-1354-2018, this FEI as the recalling firmFeb 27, 20184 x 0.251
Class II recall Z-2064-2018, this FEI as the recalling firmDec 13, 20174 x 0.251
Class II recall Z-0565-2018, this FEI as the recalling firmNov 10, 20174 x 0.251
Class II recall Z-2243-2018, this FEI as the recalling firmNov 10, 20174 x 0.251
Class II recall Z-1241-2019, this FEI as the recalling firmNov 8, 20174 x 0.251
Class II recall Z-0271-2018, this FEI as the recalling firmOct 16, 20174 x 0.251
Class II recall Z-0488-2018, this FEI as the recalling firmSep 29, 20174 x 0.251
Class II recall Z-1307-2018, this FEI as the recalling firmSep 14, 20174 x 0.251
Class II recall Z-3226-2017, this FEI as the recalling firmJul 28, 20174 x 0.251
Class II recall Z-2872-2017, this FEI as the recalling firmJun 2, 20174 x 0.251
Class II recall Z-3042-2017, this FEI as the recalling firmMar 31, 20174 x 0.251
Class II recall Z-1871-2017, this FEI as the recalling firmMar 9, 20174 x 0.251
Class II recall Z-0267-2018, this FEI as the recalling firmDec 9, 20164 x 0.251
Class II recall Z-0403-2017, this FEI as the recalling firmOct 13, 201610 years or older, so it adds no points0
Class II recall Z-0363-2017, this FEI as the recalling firmOct 12, 201610 years or older, so it adds no points0
Class II recall Z-0585-2017, this FEI as the recalling firmOct 12, 201610 years or older, so it adds no points0
Class II recall Z-0600-2017, this FEI as the recalling firmOct 5, 201610 years or older, so it adds no points0
Class II recall Z-0735-2017, this FEI as the recalling firmAug 3, 201610 years or older, so it adds no points0
Class II recall Z-0967-2017, this FEI as the recalling firmJul 11, 201610 years or older, so it adds no points0
Class II recall Z-2830-2016, this FEI as the recalling firmJul 11, 201610 years or older, so it adds no points0
Inspection classified VAIApr 19, 201610 years or older, so it adds no points0
Class II recall Z-1395-2016, this FEI as the recalling firmMar 4, 201610 years or older, so it adds no points0
Class II recall Z-1357-2016, this FEI as the recalling firmDec 18, 201510 years or older, so it adds no points0
Class II recall Z-0760-2016, this FEI as the recalling firmNov 10, 201510 years or older, so it adds no points0
Class II recall Z-1617-2016, this FEI as the recalling firmNov 9, 201510 years or older, so it adds no points0
Class II recall Z-2732-2015, this FEI as the recalling firmAug 24, 201510 years or older, so it adds no points0
Class II recall Z-2384-2015, this FEI as the recalling firmJul 7, 201510 years or older, so it adds no points0
Class II recall Z-1358-2015, this FEI as the recalling firmJan 30, 201510 years or older, so it adds no points0
Class II recall Z-0163-2015, this FEI as the recalling firmOct 7, 201410 years or older, so it adds no points0
Class II recall Z-0010-2013, this FEI as the recalling firmSep 7, 201210 years or older, so it adds no points0
Inspection classified VAIJul 25, 201210 years or older, so it adds no points0
Class II recall Z-2327-2012, this FEI as the recalling firmJun 29, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing C.R. BARD, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check C.R. BARD, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
C.R. Bard Inc
Location
COVINGTON, GA (US)
Registration Expires
Dec 31, 2027
Operations
STERILIZE
Source
both