BIOMET MICROFIXATION FDA Inspection Record
BIOMET MICROFIXATION is an FDA-registered facility (FEI 1032347) located in Jacksonville, FL, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 4, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (5 VAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 5, latest Apr 11, 2017
- Import refusals
- 4
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 4, 2025 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| May 24, 2018 | Voluntary Action Indicated (VAI) | Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Sep 1, 2017 | Voluntary Action Indicated (VAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate procedures (21 CFR 820.100(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
| Oct 30, 2015 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Jan 17, 2012 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.80(e)) Records of complaint investigation (21 CFR 820.198(e)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 11, 2017 | Class II | Z-2073-2017 | Biomet Microfixation TraumaOne System Surgical Tray - Model # 46-1600 Part of the number on the tray lid differ from the part number on the base. |
| Jun 13, 2016 | Class II | Z-2751-2016 | Hard Tissue Replacement (HTR) Implant, Rx only, for use in single procedure only. Custom… Potential sterility deficiency. |
| Aug 6, 2015 | Class II | Z-0136-2016 | Neuro Combo Set Case
The cases hold surgical instrumentation in place during handling and… and 1 more product The lids on implant caddies within P/N 17-5000 and 19-5000 could not be opened by the user without application of undesirable/unexpected… |
| Oct 17, 2014 | Class II | Z-0509-2015 | Total Temporomandibular Joint Replacement System- 45mm Lft Narrow Ti Mand and 13 more products Laser etching on the parts is wider and deeper than the conditions previously validated. |
| Sep 20, 2013 | Class II | Z-2566-2014 | Biomet Microfixation Temporomandibular Joint (TMJ) Patient Matched Left Fossa Part#… Product mix occurred prior to final package and labeling process. |
FDA record score
10 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Mar 4, 2025 | 6 x 1 | 6 |
| Inspection classified VAI | May 24, 2018 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Sep 1, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-2073-2017, this FEI as the recalling firm | Apr 11, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2751-2016, this FEI as the recalling firm | Jun 13, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 30, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0136-2016, this FEI as the recalling firm | Aug 6, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0509-2015, this FEI as the recalling firm | Oct 17, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2566-2014, this FEI as the recalling firm | Sep 20, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jan 17, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 27, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 27, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 27, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Refresh overdue |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing BIOMET MICROFIXATION as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for BIOMET MICROFIXATION: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Facility details
- Company
- Zimmer Biomet CMF And Thoracic LLC
- Location
- Jacksonville, FL (US)
- Source
- device