Wilson-Cook Medical Inc FDA Inspection Record

FEI 1037905 · Inspections, Form 483s, warning letters and import refusals

Wilson-Cook Medical Inc is an FDA-registered facility (FEI 1037905) located in Winston Salem, NC, US. Registered operations: REPACK. RegDataLab tracks 4 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Oct 24, 2025: No Action Indicated (NAI)
Inspections on record
4 (2 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Jun 24, 2011. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
14, latest Aug 20, 2026

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 24, 2025No Action Indicated (NAI)No citations listed
Dec 23, 2014Voluntary Action Indicated (VAI)Documentation (21 CFR 820.100(b))
Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2))
Mar 6, 2012Voluntary Action Indicated (VAI)Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e))
Documentation (21 CFR 820.80(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
Sampling plans (21 CFR 820.250(b))
and 2 more
Apr 19, 2011No Action Indicated (NAI)Lack of or inadequate procedures for storage. (21 CFR 820.150)
Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
and 4 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Showing the 10 most recent of 14 recalls. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 20, 2026Class IIZ-3177-2026Brand Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Product Name: HEMOSPRAY ENDOSCOPIC…
Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU).
Mar 3, 2026Class IIZ-1812-2026Instinct Plus Endoscopic Clipping Device REF: G58010 Rx Only, Sterile EO
Due to increased in complaints their is the potential for endoscopic clipping device to malfunction.
Dec 18, 2025Class IIZ-1362-2026Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI…
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an…
Dec 15, 2025Class IIZ-1298-2026Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10
Nasal feeding tube packaged without the nasal transfer tube component.
Oct 31, 2024Class IIZ-0582-2025Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking…
Aug 29, 2024Class IIZ-0036-2025Cook Medical Hemo-7 Hemospray Endoscopic Hemostat, REF G56572, 2.8 mm, Sterile; and…
Specific lots of product have been manufactured with nonconforming raw materials.
Nov 3, 2023Class IIZ-0457-2024QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3, Order Number G22657; used to dilate… and 6 more products
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon…
Jun 22, 2023Class IIZ-2252-2023Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy… and 1 more product
Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from…
May 22, 2023Class IIZ-1978-2023Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b)… and 1 more product
There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope…
Feb 4, 2020Class IIZ-1745-2020HEMO-7 Hemospray Endoscopic Hemostat and 1 more product
Complaints that the handle and/or activation knob on Hemostat devices have cracked or broken when the device is activated. In some cases…

FDA record score

33 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 28, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

17 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-3177-2026, this FEI as the recalling firmAug 20, 20264 x 14
Class II recall Z-1812-2026, this FEI as the recalling firmMar 3, 20264 x 14
Class II recall Z-1362-2026, this FEI as the recalling firmDec 18, 20254 x 14
Class II recall Z-1298-2026, this FEI as the recalling firmDec 15, 20254 x 14
Class II recall Z-0582-2025, this FEI as the recalling firmOct 31, 20244 x 0.62.4
Class II recall Z-0036-2025, this FEI as the recalling firmAug 29, 20244 x 0.62.4
Class II recall Z-0457-2024, this FEI as the recalling firmNov 3, 20234 x 0.62.4
Class II recall Z-2252-2023, this FEI as the recalling firmJun 22, 20234 x 0.62.4
Class II recall Z-1978-2023, this FEI as the recalling firmMay 22, 20234 x 0.62.4
Class II recall Z-1745-2020, this FEI as the recalling firmFeb 4, 20204 x 0.251
Class II recall Z-2541-2019, this FEI as the recalling firmJul 24, 20194 x 0.251
Class II recall Z-2089-2019, this FEI as the recalling firmJun 14, 20194 x 0.251
Class II recall Z-1132-2019, this FEI as the recalling firmJan 4, 20194 x 0.251
Class II recall Z-0774-2019, this FEI as the recalling firmDec 18, 20184 x 0.251
Inspection classified VAIDec 23, 201410 years or older, so it adds no points0
Inspection classified VAIMar 6, 201210 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsJun 24, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Wilson-Cook Medical Inc as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Wilson-Cook Medical Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Wilson-Cook Medical Inc → Check this supplier

Facility details

Company
Wilson-Cook Medical Inc.
Location
Winston Salem, NC (US)
Registration Expires
Dec 31, 2026
Operations
REPACK
Source
both

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.