ZimVie US Corp LLC FDA Inspection Record
ZimVie US Corp LLC is an FDA-registered facility (FEI 1038806) located in Palm Beach Gardens, FL, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jul 11, 2017: No Action Indicated (NAI)
- Inspections on record
- 6 (1 OAI, 3 VAI, 2 NAI)
- Last OAI inspection
- Aug 2, 2013
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 24, latest Aug 31, 2020
- Import refusals
- 4
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 11, 2017 | No Action Indicated (NAI) | No citations listed |
| Aug 6, 2014 | Voluntary Action Indicated (VAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| May 14, 2014 | No Action Indicated (NAI) | No citations listed |
| Aug 2, 2013 | Official Action Indicated (OAI) | Design input - Lack of or inadequate procedures (21 CFR 820.30(c)) Incorrect translation to production specifications (21 CFR 820.30(h)) Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) and 4 more |
| Jun 15, 2012 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate procedures (21 CFR 820.100(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) and 1 more |
| May 1, 2009 | Voluntary Action Indicated (VAI) | Incomplete, ambiguous, conflicting design requirements (21 CFR 820.30(c)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) Results of investigation (21 CFR 820.198(e)(6)) CAPA verification/validation of corrective/preventive action (21 CFR 820.100(a)(4)) and 2 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Aug 31, 2020 | Class II | Z-0073-2021 | BellaTek Encode Defin3DPMSERitive Abutments-3D Printed Model
Reference Numbers: 3DPMSER… and 6 more products Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan… |
| Jan 10, 2020 | Class II | Z-1366-2020 | CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 3.8MM(P) X 3MM(H)
Reference Number… and 22 more products Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30… |
| Feb 26, 2019 | Class II | Z-2140-2019 | CERTAIN(R) BELLATEK(R) TIN ABUTMENT 3.4MM, Item Number IEDAN3 - Product Usage: BIOMET3i… and 4 more products During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head… |
| Mar 8, 2017 | Class II | Z-1701-2017 | BIOMET 3i LTX DENTAL IMPLANTS
Model LTX610 (6x10mm) - Plastic tray within Tyvek tray… A single lot of product that potentially has a single part within the lot that does not match the product on the label due to a commingle at… |
| Sep 14, 2016 | Class II | Z-0828-2017 | BIOMET 3i CERTAIN LOW PROFILE 17¿ ABUTMENT 3.4MM(D) X 2MM(H)
An endosseous dental implant… and 35 more products A limited number of product pouches may not have been sealed during packaging. |
| Aug 8, 2016 | Class II | Z-2845-2016 | 3I OSSEOTITE CERTAIN DENTAL IMPLANTS, 4x11.5 mm, Sterile, Rx only. Product was incorrectly packaged. |
| Nov 25, 2015 | Class II | Z-2406-2016 | Certain EP Healing Abutment 5mm(D) x 5.6mm(P) x 2mm(H)
Item: IWTH52
Dental implants and 118 more products Pouches may not have been sealed during packing. |
| Nov 16, 2015 | Class II | Z-0564-2016 | Radiographic Transparency for Certain and External Connection Tapered Implants. Images… Images contained within radiographic transparency ART 1036 are undersized. |
| Sep 21, 2015 | Class II | Z-0591-2016 | BIOMET 3i, PROVIDE PROTECTION CAP, Catalog # PPC484, PPC485, PPC654, and PPC655
BIOMET… and 3 more products Inadequate biocompatibility testing. |
| Aug 10, 2015 | Class II | Z-0266-2016 | Biomet 3i Narrow Right Angle Large Driver Tip (Hexed). Driver tips are attached to a… and 1 more product The RASH3N, a hexed, abutment screw driver tip has been labeled and distributed as RASQ3N, a square, abutment screw driver tip… |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
27 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-0073-2021, this FEI as the recalling firm | Aug 31, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-1366-2020, this FEI as the recalling firm | Jan 10, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2140-2019, this FEI as the recalling firm | Feb 26, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1701-2017, this FEI as the recalling firm | Mar 8, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0828-2017, this FEI as the recalling firm | Sep 14, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2845-2016, this FEI as the recalling firm | Aug 8, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2406-2016, this FEI as the recalling firm | Nov 25, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0564-2016, this FEI as the recalling firm | Nov 16, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0591-2016, this FEI as the recalling firm | Sep 21, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0266-2016, this FEI as the recalling firm | Aug 10, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2760-2015, this FEI as the recalling firm | Jul 29, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1409-2015, this FEI as the recalling firm | Nov 7, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 6, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2594-2014, this FEI as the recalling firm | Jun 25, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1458-2014, this FEI as the recalling firm | Aug 29, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1215-2014, this FEI as the recalling firm | Aug 26, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Aug 2, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1167-2014, this FEI as the recalling firm | Jul 22, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2207-2013, this FEI as the recalling firm | Jun 10, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1960-2013, this FEI as the recalling firm | May 10, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0297-2013, this FEI as the recalling firm | Sep 28, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0119-2013, this FEI as the recalling firm | Aug 3, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 15, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2271-2012, this FEI as the recalling firm | May 31, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2158-2012, this FEI as the recalling firm | Mar 23, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1828-2012, this FEI as the recalling firm | Nov 4, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 1, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 28, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 28, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 28, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 27, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |