Bausch & Lomb Incorporated FDA Inspection Record

FEI 1119279 · Inspections, Form 483s, warning letters and import refusals

Bausch & Lomb Incorporated is an FDA-registered facility (FEI 1119279) located in CLEARWATER, FL, US. RegDataLab tracks 7 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Nov 3, 2017: No Action Indicated (NAI)
Inspections on record
7 (3 VAI, 4 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
8, latest Jun 6, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Nov 3, 2017No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
DMR - not or inadequately maintained (21 CFR 820.181)
Documentation (21 CFR 820.100(b))
Environmental control Lack of or inadequate procedures (21 CFR 820.70(c))
Apr 5, 2017Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Investigation of device failures (21 CFR 820.198(c))
Production processes (21 CFR 820.70(a))
Lack of or inadequate process validation (21 CFR 820.75(a))
Nov 30, 2015No Action Indicated (NAI)No citations listed
May 18, 2015No Action Indicated (NAI)DHR - not or inadequately maintained (21 CFR 820.184)
Nonconforming product control (21 CFR 820.90(a))
Jul 22, 2013Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a))
Documentation (21 CFR 820.100(b))
Oct 7, 2011No Action Indicated (NAI)No citations listed
Nov 23, 2009Voluntary Action Indicated (VAI)Availability (21 CFR 820.180)
DMR required activities {see also 820.181} (21 CFR 820.80(d)(1))
General (21 CFR 820.198(a))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jun 6, 2025Class IIIZ-2451-2025Akreos, SKU: AO60P0300. Akreos intraocular lenses
An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected…
Mar 27, 2025Class IZ-1679-2025Intraocular lens. enVista Monofocal IOL, All models starting with EE; enVista Aspire IOL… and 1 more product
In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire…
Nov 7, 2020Class IIZ-0519-2021MX60ET (enhanced enVista Toric) enVista¿ One-Piece Hydrophobic Acrylic Toric Model MX60ET…
Product may be missing toric axis marks.
Jun 9, 2020Class IIZ-2552-2020enVista one-piece hydrophobic acrylic toric intraocular lens, Bausch and Lomb - Product…
The toric axis marks were misaligned (from the expected position in relation to the haptics) on an implanted lens.
Nov 5, 2018Class IIZ-0628-2019enVista One Piece Hydrophobic Acrylic Intraocular lens - Is indicated for primary…
Cosmetic imperfections on the surface of some lenses.
Apr 5, 2017Class IIZ-2098-2017SoftPort Easy-Load Lens Delivery System Model: EZ-28V Is used for the folding and…
Complaints concerning kinked haptics that were occurring during lens delivery.
May 26, 2015Class IIZ-2153-2015Bausch & Lomb SofPort Advanced Optics Aspheric Lens and 1 more product
Haptics breaking during lens loading and insertion.
Oct 16, 2013Class IIZ-1062-2014Bausch & Lomb AKREOS AO Micro Incision Lens The Akreos IOL intended for primary…
Lens was manufactured with incorrect raw material.

FDA record score

21 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 29, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class I recall Z-1679-2025, this FEI as the recalling firmMar 27, 202515 x 115
Inspection classified VAIApr 5, 20176 x 0.251.5
Class II recall Z-0519-2021, this FEI as the recalling firmNov 7, 20204 x 0.251
Class II recall Z-2552-2020, this FEI as the recalling firmJun 9, 20204 x 0.251
Class II recall Z-0628-2019, this FEI as the recalling firmNov 5, 20184 x 0.251
Class II recall Z-2098-2017, this FEI as the recalling firmApr 5, 20174 x 0.251
Class II recall Z-2153-2015, this FEI as the recalling firmMay 26, 201510 years or older, so it adds no points0
Class II recall Z-1062-2014, this FEI as the recalling firmOct 16, 201310 years or older, so it adds no points0
Inspection classified VAIJul 22, 201310 years or older, so it adds no points0
Inspection classified VAINov 23, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 29, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 29, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 29, 2026Newest record Nov 19, 2026Current
Warning letter textNov 29, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 28, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 29, 2026Newest record Nov 19, 2026Current

Reviewing Bausch & Lomb Incorporated as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Bausch & Lomb Incorporated: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Bausch & Lomb Incorporated → Check this supplier

Facility details

Company
Bausch & Lomb Surgical, Inc.
Location
CLEARWATER, FL (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.