Bausch & Lomb Incorporated FDA Inspection Record
Bausch & Lomb Incorporated is an FDA-registered facility (FEI 1119279) located in CLEARWATER, FL, US. RegDataLab tracks 7 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Nov 3, 2017: No Action Indicated (NAI)
- Inspections on record
- 7 (3 VAI, 4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Jun 6, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 3, 2017 | No Action Indicated (NAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) DMR - not or inadequately maintained (21 CFR 820.181) Documentation (21 CFR 820.100(b)) Environmental control Lack of or inadequate procedures (21 CFR 820.70(c)) |
| Apr 5, 2017 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Production processes (21 CFR 820.70(a)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Nov 30, 2015 | No Action Indicated (NAI) | No citations listed |
| May 18, 2015 | No Action Indicated (NAI) | DHR - not or inadequately maintained (21 CFR 820.184) Nonconforming product control (21 CFR 820.90(a)) |
| Jul 22, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Documentation (21 CFR 820.100(b)) |
| Oct 7, 2011 | No Action Indicated (NAI) | No citations listed |
| Nov 23, 2009 | Voluntary Action Indicated (VAI) | Availability (21 CFR 820.180) DMR required activities {see also 820.181} (21 CFR 820.80(d)(1)) General (21 CFR 820.198(a)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jun 6, 2025 | Class III | Z-2451-2025 | Akreos, SKU: AO60P0300. Akreos intraocular lenses An off-power error was identified, which resulted in a diopter of 16, rather than the intended 3 diopter. Implantation of affected… |
| Mar 27, 2025 | Class I | Z-1679-2025 | Intraocular lens. enVista Monofocal IOL, All models starting with EE;
enVista Aspire IOL… and 1 more product In response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire… |
| Nov 7, 2020 | Class II | Z-0519-2021 | MX60ET (enhanced enVista Toric) enVista¿ One-Piece Hydrophobic Acrylic Toric Model MX60ET… Product may be missing toric axis marks. |
| Jun 9, 2020 | Class II | Z-2552-2020 | enVista one-piece hydrophobic acrylic toric intraocular lens, Bausch and Lomb - Product… The toric axis marks were misaligned (from the expected position in relation to the haptics) on an implanted lens. |
| Nov 5, 2018 | Class II | Z-0628-2019 | enVista One Piece Hydrophobic Acrylic Intraocular lens -
Is indicated for primary… Cosmetic imperfections on the surface of some lenses. |
| Apr 5, 2017 | Class II | Z-2098-2017 | SoftPort Easy-Load Lens Delivery System Model: EZ-28V
Is used for the folding and… Complaints concerning kinked haptics that were occurring during lens delivery. |
| May 26, 2015 | Class II | Z-2153-2015 | Bausch & Lomb SofPort Advanced Optics Aspheric Lens and 1 more product Haptics breaking during lens loading and insertion. |
| Oct 16, 2013 | Class II | Z-1062-2014 | Bausch & Lomb AKREOS AO Micro Incision Lens
The Akreos IOL intended for primary… Lens was manufactured with incorrect raw material. |
FDA record score
21 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
10 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-1679-2025, this FEI as the recalling firm | Mar 27, 2025 | 15 x 1 | 15 |
| Inspection classified VAI | Apr 5, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0519-2021, this FEI as the recalling firm | Nov 7, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-2552-2020, this FEI as the recalling firm | Jun 9, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0628-2019, this FEI as the recalling firm | Nov 5, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-2098-2017, this FEI as the recalling firm | Apr 5, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2153-2015, this FEI as the recalling firm | May 26, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1062-2014, this FEI as the recalling firm | Oct 16, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jul 22, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 23, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 29, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 29, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 29, 2026 | Newest record Nov 24, 2026 | Current |
| Recalls (Enforcement Report) | Nov 28, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 29, 2026 | Newest record Nov 19, 2026 | Current |
Reviewing Bausch & Lomb Incorporated as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Bausch & Lomb Incorporated: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Bausch & Lomb Incorporated → Check this supplier
Facility details
- Company
- Bausch & Lomb Surgical, Inc.
- Location
- CLEARWATER, FL (US)
- Source
- device