Manus Medical, LLC FDA Inspection Record

FEI 1121091 · Inspections, Form 483s, warning letters and import refusals

Manus Medical, LLC is an FDA-registered facility (FEI 1121091) located in Henrico, VA, US. RegDataLab tracks 5 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jun 26, 2024: No Action Indicated (NAI)
Inspections on record
5 (3 VAI, 2 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
4, latest Dec 20, 2019

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 26, 2024No Action Indicated (NAI)No citations listed
Oct 26, 2022Voluntary Action Indicated (VAI)Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b))
Quality Audits - defined intervals (21 CFR 820.22)
Sep 4, 2019Voluntary Action Indicated (VAI)Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Quality Audits - defined intervals (21 CFR 820.22)
Justification for not reporting (21 CFR 806.20(b)(4))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
Sep 21, 2016Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design control - no procedures (21 CFR 820.30(a))
Software validation for automated processes (21 CFR 820.70(i))
and 8 more
Nov 5, 2012No Action Indicated (NAI)Quality audits - auditor independence (21 CFR 820.22)
Documentation (21 CFR 820.100(b))

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Dec 20, 2019Class IIZ-0863-2020Custom Healthcare Systems, Inc. Inpatient Dialysis Kit PAK-19244. Single use. Dialysis…
A component (1961 TELFA PAD 2 X 3) contains LATEX and was used in the Medical Convenience Kit(PAK-19244). Case label and device label for…
Aug 13, 2018Class IIZ-0240-2019CHS Custom Convenience Kit-DR. NANDA CLOSING TRAY(CONMED HYFRECATOR SHARP) Product… and 28 more products
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
Nov 5, 2014Class IIZ-2096-2015Safety PICC Insertion Tray with Chloraprep OneStep - Model #5026CL-9 - 12 pr case…
An INCORRECT item number is packed within the OUTSIDE "PIGGYBACKED" PACKAGE. The interior sealed STERILE KIT is correct and NO adverse…
Nov 26, 2012Class IIZ-0796-2013Custom Breast Biopsy Tray, Catalog # B 9-16261, packaged in bags, 1 tray/bag, 20… and 1 more product
Surgical procedure trays are packed with pharmaceuticals for injection that may contain particulate matter.

FDA record score

7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 28, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIOct 26, 20226 x 0.63.6
Inspection classified VAISep 4, 20196 x 0.251.5
Class II recall Z-0863-2020, this FEI as the recalling firmDec 20, 20194 x 0.251
Class II recall Z-0240-2019, this FEI as the recalling firmAug 13, 20184 x 0.251
Inspection classified VAISep 21, 201610 years or older, so it adds no points0
Class II recall Z-2096-2015, this FEI as the recalling firmNov 5, 201410 years or older, so it adds no points0
Class II recall Z-0796-2013, this FEI as the recalling firmNov 26, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 27, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 27, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 27, 2026Newest record Nov 19, 2026Current
Warning letter textNov 27, 2026Newest record Nov 24, 2026Current
Recalls (Enforcement Report)Nov 27, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Manus Medical, LLC as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Manus Medical, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Manus Medical, LLC → Check this supplier

Facility details

Company
MANUS MEDICAL, LLC
Location
Henrico, VA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.