INTEGRA LIFESCIENCES CORPORATION FDA Inspection Record

FEI 1121308 · Inspections, Form 483s, warning letters and import refusals

INTEGRA LIFESCIENCES CORPORATION is an FDA-registered facility (FEI 1121308) located in PLAINSBORO, NJ, US. RegDataLab tracks 7 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Aug 22, 2024: No Action Indicated (NAI)
Inspections on record
7 (1 OAI, 2 VAI, 4 NAI)
Last OAI inspection
Aug 25, 2011
Warning letters
1 listed by FDA for this FEI, latest Dec 21, 2011
Form 483 and inspection documents
1 in FDA's reading room
Recalls as the recalling firm
6, latest Mar 11, 2016

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Aug 22, 2024No Action Indicated (NAI)No citations listed
Jun 15, 2017No Action Indicated (NAI)No citations listed
Dec 17, 2015No Action Indicated (NAI)No citations listed
Aug 14, 2013No Action Indicated (NAI)No citations listed
Jul 30, 2012Voluntary Action Indicated (VAI)Lack of or inadequate procedures for storage. (21 CFR 820.150)
Procedures not adequately established or maintained (21 CFR 820.40)
Software validation for automated processes (21 CFR 820.70(i))
Nonconforming product control (21 CFR 820.90(a))
and 3 more
Aug 25, 2011Official Action Indicated (OAI)Buildings (21 CFR 820.70(f))
Lack of or inadequate procedures (21 CFR 820.100(a))
Evaluation, timeliness, identification (21 CFR 820.198(d))
Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1))
and 4 more
Nov 20, 2009Voluntary Action Indicated (VAI)Evaluation, timeliness, identification (21 CFR 820.198(d))
Lack of or inadequate process validation (21 CFR 820.75(a))
Report of Death or Serious Injury (21 CFR 803.50(a)(1))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
and 1 more

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Mar 11, 2016Class IIZ-2777-2016Integra Meshed Dermal Regeneration Template 5 cmx 5 cm (2 in x 2in) Rx Only Meshed…
Integra's post QA release review of historical product release test results for Meshed IDRT products identified the Peel Strength test…
Mar 28, 2014Class IIZ-1697-2014Integra Newdeal SURFIX Fixed Angle Locking Screw 3.5mm x 22mm (catalog number 285322S) and…
New Deal has identified through a complaint report that some 3.5 mm Diameter x 22 mm Length Surfix screws in Lot no. F626 (Catalogue No…
May 9, 2013Class IIZ-1736-2013Integra PANTA Arthrodesis Nail System. Intended for use in tibiotalocalcaneal…
As a result of the investigation into an adverse trend of complaints for reports of misalignment of the calcaneal or tibial screws during…
Mar 28, 2013Class IIZ-1282-2013Dural Graft Matrix DuraGen¿ Dural Graft Matrix is an absorbable implant for repair of… and 20 more products
Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the…
Aug 9, 2012Class IIZ-2428-2012Integra CRW Precision Arc (CRWPRECISE), CRW Arc System (CRWASL), HRAIM Head Ring Assembly… and 1 more product
T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly…
Apr 4, 2012Class IIZ-1984-2012Pudenz Flushing Valve with Integral Connectors, catalog number NL850-1330. Integra…
The pouch label of catalog NL850-1330 indicates an incorrect model and closing pressure of the device. The incorrect label reads: "Model: 12…

Form 483s and inspection documents

  • 483 on FDA.gov
    Inspection Jul 22, 2011 to Aug 25, 2011, issued Aug 25, 2011 (dates printed on the document). FDA listing date Jul 11, 2015. Published by FDA Jan 18, 2024.

FDA record score

0 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 17, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

10 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2777-2016, this FEI as the recalling firmMar 11, 201610 years or older, so it adds no points0
Class II recall Z-1697-2014, this FEI as the recalling firmMar 28, 201410 years or older, so it adds no points0
Class II recall Z-1736-2013, this FEI as the recalling firmMay 9, 201310 years or older, so it adds no points0
Class II recall Z-1282-2013, this FEI as the recalling firmMar 28, 201310 years or older, so it adds no points0
Class II recall Z-2428-2012, this FEI as the recalling firmAug 9, 201210 years or older, so it adds no points0
Inspection classified VAIJul 30, 201210 years or older, so it adds no points0
Class II recall Z-1984-2012, this FEI as the recalling firmApr 4, 201210 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsDec 21, 201110 years or older, so it adds no points0
Inspection classified OAIAug 25, 201110 years or older, so it adds no points0
Inspection classified VAINov 20, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 16, 2026Newest record Nov 5, 2026Current
Inspection citationsNov 16, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 2, 2026Newest record Oct 29, 2026Current
Warning letter textNov 16, 2026Newest record Nov 10, 2026Current
Recalls (Enforcement Report)Nov 16, 2026Newest record Nov 11, 2026Current
Form 483s and inspection documentsNov 13, 2026Newest record Nov 12, 2026Current
Import refusalsNov 15, 2026Newest record Nov 5, 2026Current

Reviewing INTEGRA LIFESCIENCES CORPORATION as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check INTEGRA LIFESCIENCES CORPORATION yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Integra LifeSciences Corporation
Location
PLAINSBORO, NJ (US)
Source
device