DAVOL INC., SUB. C.R. BARD, INC. FDA Inspection Record
DAVOL INC., SUB. C.R. BARD, INC. is an FDA-registered facility (FEI 1213643) located in Warwick, RI, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Dec 7, 2022: No Action Indicated (NAI)
- Inspections on record
- 6 (3 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Mar 25, 2026
- Import refusals
- 2
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Dec 7, 2022 | No Action Indicated (NAI) | No citations listed |
| Feb 7, 2017 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Sep 5, 2012 | Voluntary Action Indicated (VAI) | Report of Death or Serious Injury (21 CFR 803.50(a)(1)) |
| Apr 13, 2010 | No Action Indicated (NAI) | No citations listed |
| Mar 17, 2010 | Voluntary Action Indicated (VAI) | Control/distribution procedures (21 CFR 820.160(a)) |
| May 15, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Mar 25, 2026 | Class II | Z-1896-2026 | Avitene Ultrafoam Microfibrillar Collagen Hemostat - 8cm x 6.25cm x 1cm. Catalog Number… Potential for product to contain foreign matter, confirmed to be inspect fragments. |
| Apr 25, 2023 | Class II | Z-2241-2023 | Ventralight ST 4" x 6" Ellipse with Echo PS
Catalog Number: 5955460 and 3 more products Balloon may not inflate as intended. If the balloon does not inflate during laparoscopic ventral hernia repair this may result in a delay… |
| Jun 27, 2022 | Class II | Z-1457-2022 | X-Stream Irrigation System (Catalog No. 5551000) and X-Stream Irrigation System… The device may potentially cause a fire, leading to a safety risk. |
| Jun 7, 2022 | Class II | Z-1304-2022 | HydroSurg Plus Laparoscopic Irrigator with Nezhat-Dorsey
SmokEvac Trumpet Valve, Handle… and 1 more product The product is not able to provide irrigation. |
| Oct 8, 2014 | Class II | Z-0191-2015 | Bard¿ PerFix Light Plug
The Bard¿ PerFix Light Plug is indicated for reinforcement if… The patient record peel-off label product code and size did not match the other labeling and actual product in the packaging for the Bard… |
| Apr 24, 2014 | Class II | Z-1682-2014 | Composix" LP with Echo PS" 10x13 HUXH0294
HUXL0314
Indicated for use in the… and 14 more products Pouch holding sterile inflation assembly and inflation adapter may be open or have a weak seal. |
| Jan 3, 2014 | Class II | Z-0819-2014 | Composix L/P with Echo PS 10" X 13" Product Code 0144113
The Composix L/P Mesh with… and 1 more product Product labeling does not match product configuration. |
| Sep 4, 2012 | Class II | Z-2412-2012 | Bard Ventralight ST Mesh. - Ellipse 8" x 10" (20.3 cm x 25.4 cm)
Product Code… and 1 more product Mislabeled: The 8" x 10" package contains 7" x 9" VENTRALIGHT ST Mesh product and the 7" x 9" package contains 8" x 10" VENTRALIGHT ST… |
FDA record score
13 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
11 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1896-2026, this FEI as the recalling firm | Mar 25, 2026 | 4 x 1 | 4 |
| Class II recall Z-2241-2023, this FEI as the recalling firm | Apr 25, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-1457-2022, this FEI as the recalling firm | Jun 27, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1304-2022, this FEI as the recalling firm | Jun 7, 2022 | 4 x 0.6 | 2.4 |
| Inspection classified VAI | Feb 7, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0191-2015, this FEI as the recalling firm | Oct 8, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1682-2014, this FEI as the recalling firm | Apr 24, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0819-2014, this FEI as the recalling firm | Jan 3, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Sep 5, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2412-2012, this FEI as the recalling firm | Sep 4, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 17, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing DAVOL INC., SUB. C.R. BARD, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check DAVOL INC., SUB. C.R. BARD, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Davol, Inc.
- Location
- Warwick, RI (US)
- Source
- device