Cadence Science, Inc. FDA Inspection Record
Cadence Science, Inc. is an FDA-registered facility (FEI 1213649) located in Cranston, RI, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Nov 29, 2022: No Action Indicated (NAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest May 10, 2019
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 29, 2022 | No Action Indicated (NAI) | No citations listed |
| Sep 13, 2019 | Voluntary Action Indicated (VAI) | Report of risk to health (21 CFR 806.10(a)(1)) Failure to Submit 5 Day Report (21 CFR 803.53(a)) |
| Jan 25, 2017 | Voluntary Action Indicated (VAI) | Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Lack of or inadequate DHR procedures (21 CFR 820.184) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) and 2 more |
| Feb 26, 2013 | Voluntary Action Indicated (VAI) | Investigation of device failures (21 CFR 820.198(c)) Lack of procedures, or not maintained (21 CFR 820.40) Lack of or inadequate process validation (21 CFR 820.75(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Mar 4, 2011 | Voluntary Action Indicated (VAI) | Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Review and evaluation for investigation (21 CFR 820.198(b)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) and 1 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 10, 2019 | Class II | Z-0365-2020 | Cadence Science Pressure Control Glass Syringe, Part Nos. 5338 (GTIN #816329023381); 5339… The glass syringe used with Pressure Control Syringes may potentially break during use. |
| Jul 9, 2015 | Class II | Z-2364-2015 | Perfectum Tuberculin Glass Syringe, Non sterile
Catalog numbers 5202, 5205, 5212 and 1 more product Graduation marks of the syringe are not centered with the numeric position marks on the syringe barrel. |
FDA record score
4 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
6 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Sep 13, 2019 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Jan 25, 2017 | 6 x 0.25 | 1.5 |
| Class II recall Z-0365-2020, this FEI as the recalling firm | May 10, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-2364-2015, this FEI as the recalling firm | Jul 9, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Feb 26, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 4, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Cadence Science, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Cadence Science, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Cadence Science, Inc.
- Location
- Cranston, RI (US)