MIRA, INC. FDA Inspection Record

FEI 1218813 · Inspections, Form 483s, warning letters and import refusals

MIRA, INC. is an FDA-registered facility (FEI 1218813) located in UXBRIDGE, MA, US. RegDataLab tracks 10 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Jul 23, 2025: No Action Indicated (NAI)
Inspections on record
10 (7 VAI, 3 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jul 23, 2025No Action Indicated (NAI)No citations listed
Mar 7, 2023No Action Indicated (NAI)Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)
Nov 18, 2021No Action Indicated (NAI)Document change records, maintained. (21 CFR 820.40(b))
Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Lack of or inadequate process validation (21 CFR 820.75(a))
Feb 20, 2018Voluntary Action Indicated (VAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Design history file (21 CFR 820.30(j))
Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
and 1 more
Nov 9, 2016Voluntary Action Indicated (VAI)Design verification - output does not meet input requirement (21 CFR 820.30(f))
Quality audits - auditor independence (21 CFR 820.22)
Lack of or inadequate procedures (21 CFR 820.100(a))
Design validation - simulated testing (21 CFR 820.30(g))
and 6 more
Jul 31, 2014Voluntary Action Indicated (VAI)Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Training - Lack of or inadequate procedures (21 CFR 820.25(b))
Design history file (21 CFR 820.30(j))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
and 7 more
Aug 26, 2011Voluntary Action Indicated (VAI)DHR - not or inadequately maintained (21 CFR 820.184)
Nonconforming product control (21 CFR 820.90(a))
Availability (21 CFR 820.180)
Investigation of device failures (21 CFR 820.198(c))
and 6 more
Jul 21, 2010Voluntary Action Indicated (VAI)Design control - no procedures (21 CFR 820.30(a))
Documentation (21 CFR 820.75(a))
Lack of or inadequate instructions (21 CFR 820.170(a))
DMR QA procedures and specifications (21 CFR 820.181(c))
and 19 more
Nov 5, 2009Voluntary Action Indicated (VAI)Quality audits - auditor independence (21 CFR 820.22)
Servicing - Lack of or inadequate procedures (21 CFR 820.200(a))
Analyzing service report (21 CFR 820.200(b))
Sep 15, 2009Voluntary Action Indicated (VAI)No citations listed

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIFeb 20, 20186 x 0.251.5
Inspection classified VAINov 9, 201610 years or older, so it adds no points0
Inspection classified VAIJul 31, 201410 years or older, so it adds no points0
Inspection classified VAIAug 26, 201110 years or older, so it adds no points0
Inspection classified VAIJul 21, 201010 years or older, so it adds no points0
Inspection classified VAINov 5, 200910 years or older, so it adds no points0
Inspection classified VAISep 15, 200910 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing MIRA, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check MIRA, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
MIRA, Inc.
Location
UXBRIDGE, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.