BELMONT MEDICAL TECHNOLOGIES FDA Inspection Record

FEI 1219702 · Inspections, Form 483s, warning letters and import refusals

BELMONT MEDICAL TECHNOLOGIES is an FDA-registered facility (FEI 1219702) located in BILLERICA, MA, US. RegDataLab tracks 2 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jun 21, 2013: Voluntary Action Indicated (VAI)
Inspections on record
2 (2 VAI)
Last OAI inspection
None on record
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest May 7, 2012. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
7, latest May 13, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jun 21, 2013Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Individual Report of Malfunction (21 CFR 803.50(a)(2))
Apr 27, 2011Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design validation - user needs and intended uses (21 CFR 820.30(g))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
May 13, 2025Class IIZ-1966-2025Allon 2001. Part Number: 200-00263
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that…
Mar 31, 2025Class IIZ-1674-2025Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use… and 2 more products
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during…
Mar 31, 2025Class IIZ-1678-20253.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032…
Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack…
Jan 2, 2025Class IIZ-1099-2025Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed…
Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential…
Jul 20, 2021Class IIZ-0153-2023CritiCool, thermoregulation device
Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The…
Sep 21, 2018Class IIIB-0162-2019The buddy Disposable Set, Part No. 905-00010, Part No. 905-00010P.
Belmont Instrument Corporation's buddy Disposable sets (Part No, 905-00010(P), not meeting release criteria, were distributed.
Sep 25, 2012Class IIZ-0058-2013Belmont Rapid Infuser, a Fluid Management System, Model FMS2000.
Leak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical…

FDA record score

17 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

9 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1966-2025, this FEI as the recalling firmMay 13, 20254 x 14
Class II recall Z-1674-2025, this FEI as the recalling firmMar 31, 20254 x 14
Class II recall Z-1678-2025, this FEI as the recalling firmMar 31, 20254 x 14
Class II recall Z-1099-2025, this FEI as the recalling firmJan 2, 20254 x 14
Class II recall Z-0153-2023, this FEI as the recalling firmJul 20, 20214 x 0.251
Inspection classified VAIJun 21, 201310 years or older, so it adds no points0
Class II recall Z-0058-2013, this FEI as the recalling firmSep 25, 201210 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsMay 7, 201210 years or older, so it adds no points0
Inspection classified VAIApr 27, 201110 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing BELMONT MEDICAL TECHNOLOGIES as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check BELMONT MEDICAL TECHNOLOGIES yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Belmont Instrument LLC
Location
BILLERICA, MA (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.