BELMONT MEDICAL TECHNOLOGIES FDA Inspection Record
BELMONT MEDICAL TECHNOLOGIES is an FDA-registered facility (FEI 1219702) located in BILLERICA, MA, US. RegDataLab tracks 2 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jun 21, 2013: Voluntary Action Indicated (VAI)
- Inspections on record
- 2 (2 VAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest May 7, 2012. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 7, latest May 13, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jun 21, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Individual Report of Malfunction (21 CFR 803.50(a)(2)) |
| Apr 27, 2011 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design validation - user needs and intended uses (21 CFR 820.30(g)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
Warning letters
- May 7, 2012: Warning letter listed on FDA's Data Dashboard. View compliance record (289900).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 13, 2025 | Class II | Z-1966-2025 | Allon 2001. Part Number: 200-00263 An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that… |
| Mar 31, 2025 | Class II | Z-1674-2025 | Hyperthermia Pump Procedure Kit. Model Number: 902-00045.
Sterile, single-use… and 2 more products Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during… |
| Mar 31, 2025 | Class II | Z-1678-2025 | 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032.
The P/N 903-00032… Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack… |
| Jan 2, 2025 | Class II | Z-1099-2025 | Belmont Medical 3-Spike Disposable Set used with the Belmont Rapid Infuser RI-2. Designed… Disposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential… |
| Jul 20, 2021 | Class II | Z-0153-2023 | CritiCool, thermoregulation device Premature pump failure may lead to the device displaying a HALT 4 error, from which the device cannot recover even after device reboot. The… |
| Sep 21, 2018 | Class III | B-0162-2019 | The buddy Disposable Set, Part No. 905-00010, Part No. 905-00010P. Belmont Instrument Corporation's buddy Disposable sets (Part No, 905-00010(P), not meeting release criteria, were distributed. |
| Sep 25, 2012 | Class II | Z-0058-2013 | Belmont Rapid Infuser, a Fluid Management System,
Model FMS2000. Leak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical… |
FDA record score
17 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1966-2025, this FEI as the recalling firm | May 13, 2025 | 4 x 1 | 4 |
| Class II recall Z-1674-2025, this FEI as the recalling firm | Mar 31, 2025 | 4 x 1 | 4 |
| Class II recall Z-1678-2025, this FEI as the recalling firm | Mar 31, 2025 | 4 x 1 | 4 |
| Class II recall Z-1099-2025, this FEI as the recalling firm | Jan 2, 2025 | 4 x 1 | 4 |
| Class II recall Z-0153-2023, this FEI as the recalling firm | Jul 20, 2021 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jun 21, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0058-2013, this FEI as the recalling firm | Sep 25, 2012 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | May 7, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 27, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing BELMONT MEDICAL TECHNOLOGIES as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check BELMONT MEDICAL TECHNOLOGIES yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Belmont Instrument LLC
- Location
- BILLERICA, MA (US)
- Source
- device