ARTHREX, INC. FDA Inspection Record
ARTHREX, INC. is an FDA-registered facility (FEI 1220246) located in Naples, FL, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 13, 2025: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 19, latest Nov 2, 2022
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 13, 2025 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of Written MDR Procedures (21 CFR 803.17) |
| May 5, 2023 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of Written MDR Procedures (21 CFR 803.17) |
| Aug 25, 2016 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) |
| Oct 30, 2009 | Voluntary Action Indicated (VAI) | Examination for accuracy (21 CFR 820.120(b)) Verification or validation of changes (21 CFR 820.70(b)) |
| Oct 9, 2009 | Voluntary Action Indicated (VAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Design input procedures - addressing conflicting requirement (21 CFR 820.30(c)) Design output procedures (21 CFR 820.30(d)) Design review - Lack of or inadequate procedures (21 CFR 820.30(e)) and 5 more |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Nov 2, 2022 | Class II | Z-0465-2023 | Small Joint OATS Set, REF AR-8981-06S, 6mm; and Small Joint OATS Set, REF AR-8981-08S… The devices may be packaged with the wrong reamer size. |
| Mar 25, 2022 | Class II | Z-1035-2022 | K-Wire, 1.35 mm x 170 mm Products do not meet length and diameter specifications. |
| Jun 14, 2021 | Class II | Z-2107-2021 | Suture Anchor, BioComposite SwiveLock C, 5.5 mm x 19.1 mm, Closed Eyelet, REF AR-2323BCC Not properly sterilized, because the outer pouch seal on the Tyvek header was sealed closely to the foil pouch. |
| Apr 2, 2021 | Class II | B-0429-2021 | Arthrex Angel PRP Kit, Angel System with Aspiration Kit, Angel System with Pwr Aspiration… and 1 more product Biological device ABS-10063 and ABS-10064, for which the sterility may have been compromised during the manufacturing process, were… |
| Mar 11, 2021 | Class II | Z-1380-2021 | Tibial Opening Wedge Osteotomy Plate, Titanium, 15.0 mm - Product Usage: use in treatment… It was discovered that five AR-13200ST-15.0 parts were processed as AR-13200T-15.0 parts. |
| Feb 24, 2021 | Class II | Z-1310-2021 | Drill Bit, Cannulated, 5.0 mm - Product Usage: may be used in a variety of foot and ankle… The color indicator on the proximal end of the shaft of the AR-8770-02 is the incorrect color. |
| Jun 24, 2020 | Class II | Z-2733-2020 | Sterile Low Profile Screw, Titanium, 6.7 x 65mm
Part Number: AR-8967-2865S - Product… Screws provided in packaging are 5mm shorter than intended |
| Dec 17, 2019 | Class II | Z-0919-2020 | Arthrex Fibulock Fibular Nail Instrument Set
Reusable non-sterile instruments
Outrigger… There is a potential for blockage of the Hub Attachment Tube. |
| Oct 1, 2019 | Class II | Z-1510-2020 | Low Profile Screw, 4.5 x 20 mm
The Low Profile ScrewTM, 4.5 x 20 mm, AR-8545-20, is… Screws provided in the device package are 6 mm longer than intended. |
| Jan 14, 2019 | Class II | Z-1206-2019 | Arthrex Burr, Oval, 12 Flute, 5.5 mm x 13 cm (AR-8550OBT)
These devices are indicated… Devices may generate excessive heat during use. |
FDA record score
27 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
22 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Jan 13, 2025 | 6 x 1 | 6 |
| Inspection classified VAI | May 5, 2023 | 6 x 0.6 | 3.6 |
| Class II recall Z-0465-2023, this FEI as the recalling firm | Nov 2, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-1035-2022, this FEI as the recalling firm | Mar 25, 2022 | 4 x 0.6 | 2.4 |
| Class II recall Z-2107-2021, this FEI as the recalling firm | Jun 14, 2021 | 4 x 0.25 | 1 |
| Class II recall B-0429-2021, this FEI as the recalling firm | Apr 2, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1380-2021, this FEI as the recalling firm | Mar 11, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1310-2021, this FEI as the recalling firm | Feb 24, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2733-2020, this FEI as the recalling firm | Jun 24, 2020 | 4 x 0.25 | 1 |
| Class II recall Z-0919-2020, this FEI as the recalling firm | Dec 17, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1510-2020, this FEI as the recalling firm | Oct 1, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-1206-2019, this FEI as the recalling firm | Jan 14, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0724-2019, this FEI as the recalling firm | Nov 19, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0442-2019, this FEI as the recalling firm | Sep 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0295-2019, this FEI as the recalling firm | Aug 27, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0095-2018, this FEI as the recalling firm | Oct 3, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2041-2017, this FEI as the recalling firm | Mar 24, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-1958-2016, this FEI as the recalling firm | Apr 22, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1360-2016, this FEI as the recalling firm | Jan 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1195-2016, this FEI as the recalling firm | Dec 17, 2015 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 30, 2009 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 9, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing ARTHREX, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check ARTHREX, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Arthrex, Inc.
- Location
- Naples, FL (US)
- Source
- device