MEDTRONIC VASCULAR FDA Inspection Record
MEDTRONIC VASCULAR is an FDA-registered facility (FEI 1220452) located in DANVERS, MA, US. RegDataLab tracks 5 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Oct 15, 2021: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (2 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 9, latest May 4, 2021
- Import refusals
- 1
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Oct 15, 2021 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Jun 2, 2016 | No Action Indicated (NAI) | No citations listed |
| Feb 27, 2013 | No Action Indicated (NAI) | No citations listed |
| Feb 24, 2011 | No Action Indicated (NAI) | No citations listed |
| Oct 30, 2008 | Voluntary Action Indicated (VAI) | Changes, communication (21 CFR 820.40(b)) Documentation of disposition, justification, signature (21 CFR 820.90(b)(1)) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| May 4, 2021 | Class I | Z-1860-2021 | Angiographic Guidewire Component: Model Number: 107042, 107044, 107447, 110003, 110004… The component part numbers intended for shipment to a third-party, where they are further processed into kits and sterilized were instead… |
| Jan 15, 2021 | Class II | Z-1245-2021 | Medtronic 6F Launcher Guide Catheter 100 cmJL 3.5
Model Number: LA6JL35 and 3 more products Sterility may be compromised due to an unsealed pouch |
| Aug 25, 2020 | Class I | Z-0087-2021 | Rashkind Balloon Septostomy Catheter, 6F, Recess Balloon, Closed end, 50 cm, Pediatric… and 2 more products Subsequent failure of catheters in the field and failed quality testing |
| Mar 15, 2019 | Class I | Z-2562-2019 | SHERPA NX ACTIVE GUIDING CATHETER, 6F, IMA 55CM .070", REF SA6IMAK. for cardiovascular… and 190 more products There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa… |
| Aug 16, 2018 | Class III | Z-3253-2018 | Medtronic DxTerity(TM) TRA 5F, Diagnostic Catheter, REF U5TRAN35 Medtronic has determined that the French size indicator on the inner pouch may incorrectly reflect a 6 French size, rather than the correct… |
| Mar 13, 2018 | Class II | Z-1479-2018 | Medtronic DxTerity Diagnostic Catheter, JL 4.0, REF DXT5JL40, marked with 6F color code on… The color coded French size indicator on the outer box may incorrectly reflect a 6 French size, rather than the correct 5 French size. The… |
| Oct 13, 2014 | Class II | Z-0205-2015 | Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K Two reports have been received on the 6F launcher Guide Catheter lot (Model Number LA6JR40K; lot No 0007244836} for incorrect usable length… |
| Feb 6, 2014 | Class III | Z-1121-2014 | Medtronic Everest 20cc Inflation Device; Model Number and Catalog Number: AC2205P
The… Incorrect master label. |
| Oct 21, 2013 | Class I | Z-0261-2014 | Cougar¿ Nitinol Workhorse Guidewire: Models: CGRLS190HJ, CGRLS190HS, CGRLS190S… and 7 more products Potential for PTFE (polytetrafluroethylene) coating to delaminate and detach from Guidewire. |
FDA record score
15 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
9 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class I recall Z-1860-2021, this FEI as the recalling firm | May 4, 2021 | 15 x 0.25 | 3.8 |
| Class I recall Z-0087-2021, this FEI as the recalling firm | Aug 25, 2020 | 15 x 0.25 | 3.8 |
| Class I recall Z-2562-2019, this FEI as the recalling firm | Mar 15, 2019 | 15 x 0.25 | 3.8 |
| Inspection classified VAI | Oct 15, 2021 | 6 x 0.25 | 1.5 |
| Class II recall Z-1245-2021, this FEI as the recalling firm | Jan 15, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1479-2018, this FEI as the recalling firm | Mar 13, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-0205-2015, this FEI as the recalling firm | Oct 13, 2014 | 10 years or older, so it adds no points | 0 |
| Class I recall Z-0261-2014, this FEI as the recalling firm | Oct 21, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 30, 2008 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |