KARL STORZ ENDOVISION, INC. FDA Inspection Record
KARL STORZ ENDOVISION, INC. is an FDA-registered facility (FEI 1221826) located in CHARLTON, MA, US. RegDataLab tracks 9 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Nov 4, 2026: Voluntary Action Indicated (VAI)
- Inspections on record
- 9 (2 OAI, 4 VAI, 3 NAI)
- Last OAI inspection
- Oct 22, 2015
- Warning letters
- 1 listed by FDA for this FEI, latest Dec 8, 2014
- Recalls as the recalling firm
- 1, latest Feb 29, 2016
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Nov 4, 2026 | Voluntary Action Indicated (VAI) | Design and development validation not performed as planned (ISO 13485:2016 Clause 7.3.7) Production and service provision not adequate (ISO 13485:2016 Clause 7.5.1) Information gathered in feedback process not used as input into risk management (ISO 13485:2016 Clause 8.2.1) Corrective action not taken (ISO 13485:2016 Clause 8.5.2) |
| Jun 14, 2022 | No Action Indicated (NAI) | Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Sep 17, 2021 | Voluntary Action Indicated (VAI) | Design verification - output does not meet input requirement (21 CFR 820.30(f)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Design history file (21 CFR 820.30(j)) |
| Jun 14, 2018 | No Action Indicated (NAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Lack of or inadequate final acceptance procedures (21 CFR 820.80(d)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) |
| Dec 20, 2016 | Voluntary Action Indicated (VAI) | Acceptance status (21 CFR 820.86) |
| Oct 22, 2015 | Official Action Indicated (OAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) and 4 more |
| Oct 29, 2014 | Official Action Indicated (OAI) | Training - Lack of or inadequate procedures (21 CFR 820.25(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) Design validation - user needs and intended uses (21 CFR 820.30(g)) and 2 more |
| Aug 26, 2013 | Voluntary Action Indicated (VAI) | DHR content (21 CFR 820.184) Maintenance schedule, Lack of or inadequate schedule (21 CFR 820.70(g)(1)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) |
| May 13, 2011 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 29, 2016 | Class II | Z-1333-2016 | ENDOSCOPE BRUSH- KARL STORZ Flexible Cleaning Brush single-use, non-sterile device for… Endoscope cleaning brush 11276CL may not fit resulting in an inability to insert the brush in to the channel of the intended endoscope or… |
FDA record score
9 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Inspection classified VAI | Nov 4, 2026 | 6 x 1 | 6 |
| Inspection classified VAI | Sep 17, 2021 | 6 x 0.25 | 1.5 |
| Inspection classified VAI | Dec 20, 2016 | 6 x 0.25 | 1.5 |
| Class II recall Z-1333-2016, this FEI as the recalling firm | Feb 29, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 22, 2015 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Dec 8, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified OAI | Oct 29, 2014 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Aug 26, 2013 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 24, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 24, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 24, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing KARL STORZ ENDOVISION, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check KARL STORZ ENDOVISION, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Karl Storz Endovision, Inc.
- Location
- CHARLTON, MA (US)
- Source
- device