DePuy Mitek, Inc. FDA Inspection Record
DePuy Mitek, Inc. is an FDA-registered facility (FEI 1221934) located in RAYNHAM, MA, US. RegDataLab tracks 2 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Mar 6, 2013: No Action Indicated (NAI)
- Inspections on record
- 2 (1 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 16, latest Feb 24, 2025
- Import refusals
- 5
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Mar 6, 2013 | No Action Indicated (NAI) | No citations listed |
| Oct 22, 2009 | Voluntary Action Indicated (VAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 24, 2025 | Class II | Z-1314-2025 | COR Disposable Kit, 8 mm. Cartilage Transplant System. and 1 more product Device is missing the pin in the graft loader component. |
| Apr 13, 2023 | Class II | Z-1677-2023 | DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone… and 4 more products IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS¿ Anchor Inserter may cause a fragment of the… |
| Nov 4, 2019 | Class II | Z-0665-2020 | Healix Knotless ADV BR 4.75 Suture Anchor and 1 more product Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead… |
| Aug 30, 2018 | Class II | Z-0208-2019 | Mitek Screw and Washer Depth Gauge, Product Code 219024 The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole. |
| Dec 22, 2017 | Class II | Z-1749-2018 | LATARJET EXPERIENCE Sharp Curved Ostetome
Product Usage:
The Latarjet EXPERIENCE… Reports of product loosening or coming unscrewed from the handle. |
| Aug 31, 2017 | Class II | Z-0695-2018 | 1113 Gravity Tube Set Pk/24 -SH
Product Usage:
The FMS Gravity Tube Set is intended to… Leakage of FMS gravity tub sets (281113) at the bulb and tube connection |
| May 12, 2017 | Class II | Z-2615-2017 | DePuy Mitek LATARJET EXPERIENCE-Combo Screw Driver
Product Code: 288211
Product… and 4 more products Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to… |
| Nov 23, 2015 | Class II | Z-0479-2016 | VAPR¿ TRIPOLAR 90" Degree Suction Electrodes (Model/Catalog No: 225028) Mitek identified that on VAPR Tripolar 90 Degree Suction Electrode the ablation and coagulation buttons are colored incorrectly. The wiring… |
| Nov 20, 2015 | Class II | Z-0656-2016 | FMS neXtra¿ 3.5mm Full Radius Cutter Shaver Blades; Product Code: 283305
Intended to… and 2 more products Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label. |
| Aug 29, 2013 | Class II | Z-0191-2014 | DePuy Mitek FMS Intermediary Tubing with One-Way valve
Product Code: 281142
Product… and 1 more product Pillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in… |
FDA record score
11 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
17 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1314-2025, this FEI as the recalling firm | Feb 24, 2025 | 4 x 1 | 4 |
| Class II recall Z-1677-2023, this FEI as the recalling firm | Apr 13, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0665-2020, this FEI as the recalling firm | Nov 4, 2019 | 4 x 0.25 | 1 |
| Class II recall Z-0208-2019, this FEI as the recalling firm | Aug 30, 2018 | 4 x 0.25 | 1 |
| Class II recall Z-1749-2018, this FEI as the recalling firm | Dec 22, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0695-2018, this FEI as the recalling firm | Aug 31, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-2615-2017, this FEI as the recalling firm | May 12, 2017 | 4 x 0.25 | 1 |
| Class II recall Z-0479-2016, this FEI as the recalling firm | Nov 23, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0656-2016, this FEI as the recalling firm | Nov 20, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0191-2014, this FEI as the recalling firm | Aug 29, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1278-2013, this FEI as the recalling firm | Apr 15, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1172-2013, this FEI as the recalling firm | Mar 22, 2013 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0529-2013, this FEI as the recalling firm | Nov 16, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0140-2013, this FEI as the recalling firm | Sep 6, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1937-2012, this FEI as the recalling firm | May 4, 2012 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0133-2013, this FEI as the recalling firm | Mar 12, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 22, 2009 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Inspection citations | Nov 14, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 2, 2026 | Newest record Oct 29, 2026 | Current |
| Warning letter text | Nov 14, 2026 | Newest record Nov 10, 2026 | Current |
| Recalls (Enforcement Report) | Nov 14, 2026 | Newest record Nov 11, 2026 | Current |
| Form 483s and inspection documents | Nov 13, 2026 | Newest record Nov 12, 2026 | Current |
| Import refusals | Nov 2, 2026 | Newest record Oct 22, 2026 | Current |
Reviewing DePuy Mitek, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check DePuy Mitek, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- DePuy Mitek, Inc., A Johnson & Johnson Co.
- Location
- RAYNHAM, MA (US)
- Source
- device