IVOCLAR VIVADENT, INC. FDA Inspection Record
IVOCLAR VIVADENT, INC. is an FDA-registered facility (FEI 1316092) located in AMHERST, NY, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Jan 11, 2018: No Action Indicated (NAI)
- Inspections on record
- 6 (3 VAI, 3 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 9, latest Dec 21, 2023
- Import refusals
- 3
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jan 11, 2018 | No Action Indicated (NAI) | Lack of or inadequate procedures (21 CFR 820.100(a)) Remedial action (21 CFR 820.72(b)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) |
| Jun 7, 2013 | Voluntary Action Indicated (VAI) | Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3)) Lack of or inadequate procedures (21 CFR 820.100(a)) Design validation acceptance criteria (21 CFR 820.30(g)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) and 1 more |
| Dec 2, 2011 | Voluntary Action Indicated (VAI) | Design verification - Lack of or inadequate procedures (21 CFR 820.30(f)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design validation acceptance criteria (21 CFR 820.30(g)) Design validation - production units (21 CFR 820.30(g)) |
| Nov 16, 2011 | Voluntary Action Indicated (VAI) | Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Investigation of device failures (21 CFR 820.198(c)) and 3 more |
| Feb 9, 2011 | No Action Indicated (NAI) | Software validation for automated processes (21 CFR 820.70(i)) Lack of or inadequate procedures (21 CFR 820.100(a)) Nonconforming product, Lack of or inadequate procedures (21 CFR 820.90(a)) Records of complaint investigation (21 CFR 820.198(e)) and 1 more |
| Mar 9, 2010 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Dec 21, 2023 | Class II | Z-1006-2024 | Article No. NA6050411, Valiant PhD No. 1 Econo Affected product may not perform as expected; the amalgam powder may not triturate correctly and the physical properties (i.e. strength… |
| Aug 3, 2017 | Class III | Z-0665-2018 | Heliomolar Refill 20x0.25g 110/A1, Product code 541501AN, Catalog number 541501 and 3 more products Complaints were filed claiming the material consistency was too soft and too sticky in comparison to its usual consistency, causing… |
| Nov 9, 2016 | Class II | Z-0717-2017 | Virtual XD Refill Light Body Fast Set Wash Material, 2X50 ml, REF #/Product Code 646461… and 2 more products The firm received complaints claiming the dental material failed to set up. As the dental material ages, the set time may increase. |
| Jul 11, 2016 | Class II | Z-2373-2016 | OptraPol Polishing Discs, Rx ONLY: OptraPol Refill Cup /10: REF/Product ID 634459… After a few seconds of polishing with moderate force, parts of the OptraPol cup are breaking away. OptraPol Small Flame is also affected… |
| May 27, 2016 | Class II | Z-0017-2017 | Zenostar MT Color Impulse blue, Art. no. 681165
Product Usage:
For coloring dental… and 20 more products The primary packaging of these products may show leaks in individual cases. As a consequence, the liquid may escape inadvertently and direct… |
| Feb 12, 2016 | Class II | Z-1026-2016 | Helioseal Clear Refill 1 x 1.25 g, REF/Product Code/Art. No. 558521AN and Helioseal Clear… Complaints were received claiming the material failed to polymerize under the curing light. After investigation, it was determined that a… |
| Jul 9, 2015 | Class II | Z-2265-2015 | IPS Empress Universal Glaze Spray 270 mL, Product Code/Model/Catalog Number 609432AN, Rx… Complaints were received of coarse particles in the glazed surface of IPS e.max Ceram restorations. A basic component of the glass contains… |
| Apr 14, 2014 | Class II | Z-1740-2014 | bluephase style Light probe Pin-point 6>2 mm black REF # 636 241. The accessory is a pin… A stock check revealed broken glass fibers between the connector and the fiber bundle in a large number of these light probes. |
| Sep 23, 2013 | Class II | Z-0376-2014 | ExciTE F DSC Soft touch Single Dose, Dual-curing Regular Ref. 50
dental adhesive for… Ivoclar Vivadent, Inc. is recalling a lot of ExciTE F DSC Dental Adhesive found to produce a blue green discoloration when used in… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
11 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1006-2024, this FEI as the recalling firm | Dec 21, 2023 | 4 x 0.6 | 2.4 |
| Class II recall Z-0717-2017, this FEI as the recalling firm | Nov 9, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2373-2016, this FEI as the recalling firm | Jul 11, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0017-2017, this FEI as the recalling firm | May 27, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1026-2016, this FEI as the recalling firm | Feb 12, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-2265-2015, this FEI as the recalling firm | Jul 9, 2015 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-1740-2014, this FEI as the recalling firm | Apr 14, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0376-2014, this FEI as the recalling firm | Sep 23, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Jun 7, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Dec 2, 2011 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 16, 2011 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 26, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 26, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 26, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 26, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing IVOCLAR VIVADENT, INC. as a supplier? Ask Claude with the RegDataLab connector to get every site's inspections, Form 483s, warning letters, recalls and import alerts, with FDA sources and dates. Prefer a PDF? Request a free supplier snapshot (up to five suppliers, prepared by hand).
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for IVOCLAR VIVADENT, INC.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about IVOCLAR VIVADENT, INC. → Check this supplier
Facility details
- Company
- Ivoclar Vivadent, Inc.
- Location
- AMHERST, NY (US)
- Source
- device