EPIMED INTERNATIONAL, INC. FDA Inspection Record
EPIMED INTERNATIONAL, INC. is an FDA-registered facility (FEI 1316297) located in JOHNSTOWN, NY, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.
What FDA found at this site
- Latest inspection
- Jul 12, 2016: Voluntary Action Indicated (VAI)
- Inspections on record
- 5 (4 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- 1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Sep 19, 2012. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 3, latest Feb 1, 2021
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Jul 12, 2016 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Design plans - Lack of or inadequate (21 CFR 820.30(b)) Design input - documentation (21 CFR 820.30(c)) Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) and 9 more |
| Nov 6, 2013 | Voluntary Action Indicated (VAI) | Design validation- Lack of or inadequate procedures (21 CFR 820.30(g)) Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i)) Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a)) |
| Mar 26, 2013 | Voluntary Action Indicated (VAI) | Lack of or inadequate complaint procedures (21 CFR 820.198(a)) Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b)) Lack of or inadequate procedures (21 CFR 820.100(a)) Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a)) and 1 more |
| May 17, 2012 | No Action Indicated (NAI) | No citations listed |
| Apr 13, 2010 | Voluntary Action Indicated (VAI) | Documentation (21 CFR 820.100(b)) Review of sampling methods for adequacy (21 CFR 820.250(b)) Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2)) Production processes (21 CFR 820.70(a)) and 2 more |
Warning letters
- Sep 19, 2012: Warning letter listed on FDA's Data Dashboard. View compliance record (317898).
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Feb 1, 2021 | Class II | Z-1276-2021 | Feth-R-Kath Catheter, 19g x 24". Intended Use:Radlo Opaque Epidural Catheter
Catalog #… Outer coating of the catheter may deteriorate and result in the coating to crack and/or potentially flake off |
| Jun 18, 2019 | Class II | Z-2068-2019 | Versa - Kath Mini Kit, with plastic L.O.R Syringe 12g X 12in (30.5 cm), 5 units, Sterile… and 1 more product Mislabeled - The product's labeling contains an incorrect size for the catheter included in the kit. The kit contains a 21 gauge catheter… |
| Apr 22, 2016 | Class II | Z-0086-2017 | 15 g R.K. Epidural Needle (TW), Catalog #/REF #100-1415
Intended for the epidural… There is a possibility that the needles may have the potential to skive a catheters outer coating. If this were to occur, there exists the… |
FDA record score
2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
8 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1276-2021, this FEI as the recalling firm | Feb 1, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-2068-2019, this FEI as the recalling firm | Jun 18, 2019 | 4 x 0.25 | 1 |
| Inspection classified VAI | Jul 12, 2016 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0086-2017, this FEI as the recalling firm | Apr 22, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Nov 6, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Mar 26, 2013 | 10 years or older, so it adds no points | 0 |
| Warning letter on FDA's dashboard, which does not show close-outs | Sep 19, 2012 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Apr 13, 2010 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 25, 2026 | Newest record Nov 20, 2026 | Current |
| Inspection citations | Nov 25, 2026 | Newest record Nov 17, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 25, 2026 | Newest record Nov 19, 2026 | Current |
| Warning letter text | Nov 24, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 25, 2026 | Newest record Nov 25, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing EPIMED INTERNATIONAL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check EPIMED INTERNATIONAL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Epimed International, Inc.
- Location
- JOHNSTOWN, NY (US)
- Source
- device