EPIMED INTERNATIONAL, INC. FDA Inspection Record

FEI 1316297 · Inspections, Form 483s, warning letters and import refusals

EPIMED INTERNATIONAL, INC. is an FDA-registered facility (FEI 1316297) located in JOHNSTOWN, NY, US. RegDataLab tracks 5 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Jul 12, 2016: Voluntary Action Indicated (VAI)
Inspections on record
5 (4 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Sep 19, 2012. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
3, latest Feb 1, 2021

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Jul 12, 2016Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design plans - Lack of or inadequate (21 CFR 820.30(b))
Design input - documentation (21 CFR 820.30(c))
Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
and 9 more
Nov 6, 2013Voluntary Action Indicated (VAI)Design validation- Lack of or inadequate procedures (21 CFR 820.30(g))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))
Mar 26, 2013Voluntary Action Indicated (VAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Lack of or inadequate receiving acceptance procedures (21 CFR 820.80(b))
Lack of or inadequate procedures (21 CFR 820.100(a))
Calibration, Inspection, etc. Procedures Lack of or Inadequ (21 CFR 820.72(a))
and 1 more
May 17, 2012No Action Indicated (NAI)No citations listed
Apr 13, 2010Voluntary Action Indicated (VAI)Documentation (21 CFR 820.100(b))
Review of sampling methods for adequacy (21 CFR 820.250(b))
Report of violation of the Act (see 803.52(e)(9)) (21 CFR 806.10(a)(2))
Production processes (21 CFR 820.70(a))
and 2 more

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Feb 1, 2021Class IIZ-1276-2021Feth-R-Kath Catheter, 19g x 24". Intended Use:Radlo Opaque Epidural Catheter Catalog #…
Outer coating of the catheter may deteriorate and result in the coating to crack and/or potentially flake off
Jun 18, 2019Class IIZ-2068-2019Versa - Kath Mini Kit, with plastic L.O.R Syringe 12g X 12in (30.5 cm), 5 units, Sterile… and 1 more product
Mislabeled - The product's labeling contains an incorrect size for the catheter included in the kit. The kit contains a 21 gauge catheter…
Apr 22, 2016Class IIZ-0086-201715 g R.K. Epidural Needle (TW), Catalog #/REF #100-1415 Intended for the epidural…
There is a possibility that the needles may have the potential to skive a catheters outer coating. If this were to occur, there exists the…

FDA record score

2 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

8 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-1276-2021, this FEI as the recalling firmFeb 1, 20214 x 0.251
Class II recall Z-2068-2019, this FEI as the recalling firmJun 18, 20194 x 0.251
Inspection classified VAIJul 12, 201610 years or older, so it adds no points0
Class II recall Z-0086-2017, this FEI as the recalling firmApr 22, 201610 years or older, so it adds no points0
Inspection classified VAINov 6, 201310 years or older, so it adds no points0
Inspection classified VAIMar 26, 201310 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsSep 19, 201210 years or older, so it adds no points0
Inspection classified VAIApr 13, 201010 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing EPIMED INTERNATIONAL, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check EPIMED INTERNATIONAL, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Epimed International, Inc.
Location
JOHNSTOWN, NY (US)
Source