Trinity Biotech USA FDA Inspection Record
Trinity Biotech USA is an FDA-registered facility (FEI 1318354) located in Jamestown, NY, US. RegDataLab tracks 3 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Sep 1, 2017: No Action Indicated (NAI)
- Inspections on record
- 3 (2 VAI, 1 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 8, latest Jul 22, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Sep 1, 2017 | No Action Indicated (NAI) | Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g)) |
| Feb 7, 2014 | Voluntary Action Indicated (VAI) | Design review - Lack of or inadequate procedures (21 CFR 820.30(e)) Lack of or inadequate procedures (21 CFR 820.100(a)) Process changes - review, evaluation and revalidation (21 CFR 820.75(c)) Documentation - review in response to changes or deviations (21 CFR 820.75(c)) and 3 more |
| May 4, 2012 | Voluntary Action Indicated (VAI) | Report of risk to health (21 CFR 806.10(a)(1)) Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50) |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Jul 22, 2025 | Class II | Z-2292-2025 | Bartels ELISA Legionella Urinary Antigen. Model Number: B1029-440.
In Vitro Diagnostic… Potential for Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 065 and 066 to contain particulate matter. |
| Oct 11, 2021 | Class II | Z-0382-2022 | Bartels Legionella Urinary Antigen- an adjunct to culture for the presumptive diagnosis of… Contamination of the conjugate may cause a pipetting issue which then could lead to an invalid assay |
| Mar 17, 2021 | Class II | Z-1503-2021 | Trinity Biotech Captia Measles IgM Kit, Catalog No. 2326060 - Product Usage: intended for… Positive Control may run high out of its assigned ISR range and result in an invalid run and necessitate retesting |
| Nov 1, 2018 | Class II | Z-0620-2019 | ultra2 Genesys Variants Analyzer, REF 03-01-0045 intended to be used with automated high… and 4 more products lack of 510K |
| Jul 11, 2016 | Class III | Z-2782-2016 | Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96… The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive… |
| Nov 17, 2014 | Class III | Z-0819-2015 | Trinity biotech Captia (TM) Syphilis IgM Capture Test Kit, Product No./REF 800-935, Test… Through in-house routine real time stability monitoring of Trinity's Captia Syphilis IgM Capture Test Kit, Lot 033, the firm identified a… |
| Aug 27, 2013 | Class II | Z-0572-2014 | Captia Syphilis-G Test Kit
Product Usage: This is an enzyme immunoassay for the… Trinity Biotech recalled Trinity Biotech Captia Syphilis IgG kits due to false negative results. |
| May 29, 2012 | Class III | Z-2445-2012 | Trinity Biotech, Captia VZV IgG. Product code 2325600: 96 Tests; product code: 2325601… The Positive Control for the Captia Varicella-Zoster Virus (VZV) IgG Kit has been reported as running high out of its assigned ISR range. |
FDA record score
7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
7 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-2292-2025, this FEI as the recalling firm | Jul 22, 2025 | 4 x 1 | 4 |
| Class II recall Z-0382-2022, this FEI as the recalling firm | Oct 11, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-1503-2021, this FEI as the recalling firm | Mar 17, 2021 | 4 x 0.25 | 1 |
| Class II recall Z-0620-2019, this FEI as the recalling firm | Nov 1, 2018 | 4 x 0.25 | 1 |
| Inspection classified VAI | Feb 7, 2014 | 10 years or older, so it adds no points | 0 |
| Class II recall Z-0572-2014, this FEI as the recalling firm | Aug 27, 2013 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | May 4, 2012 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 23, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 23, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 23, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 23, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 23, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing Trinity Biotech USA as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check Trinity Biotech USA yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Trinity Biotech USA
- Location
- Jamestown, NY (US)
- Source
- device