Trinity Biotech USA FDA Inspection Record

FEI 1318354 · Inspections, Form 483s, warning letters and import refusals

Trinity Biotech USA is an FDA-registered facility (FEI 1318354) located in Jamestown, NY, US. RegDataLab tracks 3 FDA inspections at this facility.

What FDA found at this site

Latest inspection
Sep 1, 2017: No Action Indicated (NAI)
Inspections on record
3 (2 VAI, 1 NAI)
Last OAI inspection
None on record
Warning letters
None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
Recalls as the recalling firm
8, latest Jul 22, 2025

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Sep 1, 2017No Action Indicated (NAI)Design Validation - Risk analysis not performed/inadequate (21 CFR 820.30(g))
Feb 7, 2014Voluntary Action Indicated (VAI)Design review - Lack of or inadequate procedures (21 CFR 820.30(e))
Lack of or inadequate procedures (21 CFR 820.100(a))
Process changes - review, evaluation and revalidation (21 CFR 820.75(c))
Documentation - review in response to changes or deviations (21 CFR 820.75(c))
and 3 more
May 4, 2012Voluntary Action Indicated (VAI)Report of risk to health (21 CFR 806.10(a)(1))
Purchasing controls, Lack of or inadequate procedures (21 CFR 820.50)

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Jul 22, 2025Class IIZ-2292-2025Bartels ELISA Legionella Urinary Antigen. Model Number: B1029-440. In Vitro Diagnostic…
Potential for Bartels ELISA Legionella Urinary Antigen kit B1029-440 lot 065 and 066 to contain particulate matter.
Oct 11, 2021Class IIZ-0382-2022Bartels Legionella Urinary Antigen- an adjunct to culture for the presumptive diagnosis of…
Contamination of the conjugate may cause a pipetting issue which then could lead to an invalid assay
Mar 17, 2021Class IIZ-1503-2021Trinity Biotech Captia Measles IgM Kit, Catalog No. 2326060 - Product Usage: intended for…
Positive Control may run high out of its assigned ISR range and result in an invalid run and necessitate retesting
Nov 1, 2018Class IIZ-0620-2019ultra2 Genesys Variants Analyzer, REF 03-01-0045 intended to be used with automated high… and 4 more products
lack of 510K
Jul 11, 2016Class IIIZ-2782-2016Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96…
The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive…
Nov 17, 2014Class IIIZ-0819-2015Trinity biotech Captia (TM) Syphilis IgM Capture Test Kit, Product No./REF 800-935, Test…
Through in-house routine real time stability monitoring of Trinity's Captia Syphilis IgM Capture Test Kit, Lot 033, the firm identified a…
Aug 27, 2013Class IIZ-0572-2014Captia Syphilis-G Test Kit Product Usage: This is an enzyme immunoassay for the…
Trinity Biotech recalled Trinity Biotech Captia Syphilis IgG kits due to false negative results.
May 29, 2012Class IIIZ-2445-2012Trinity Biotech, Captia VZV IgG. Product code 2325600: 96 Tests; product code: 2325601…
The Positive Control for the Captia Varicella-Zoster Virus (VZV) IgG Kit has been reported as running high out of its assigned ISR range.

FDA record score

7 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 23, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

7 records behind this number
RecordDatePoints x age weightAdds
Class II recall Z-2292-2025, this FEI as the recalling firmJul 22, 20254 x 14
Class II recall Z-0382-2022, this FEI as the recalling firmOct 11, 20214 x 0.251
Class II recall Z-1503-2021, this FEI as the recalling firmMar 17, 20214 x 0.251
Class II recall Z-0620-2019, this FEI as the recalling firmNov 1, 20184 x 0.251
Inspection classified VAIFeb 7, 201410 years or older, so it adds no points0
Class II recall Z-0572-2014, this FEI as the recalling firmAug 27, 201310 years or older, so it adds no points0
Inspection classified VAIMay 4, 201210 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 23, 2026Newest record Nov 9, 2026Current
Inspection citationsNov 23, 2026Newest record Nov 5, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 23, 2026Newest record Nov 13, 2026Current
Warning letter textNov 23, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 23, 2026Newest record Nov 18, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Trinity Biotech USA as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Trinity Biotech USA yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Trinity Biotech USA
Location
Jamestown, NY (US)
Source
device

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.