Repro-Med Systems, Inc. FDA Inspection Record

FEI 1318360 · Inspections, Form 483s, warning letters and import refusals

Repro-Med Systems, Inc. is an FDA-inspected facility (FEI 1318360) located in Chester, NY, US. RegDataLab tracks 6 FDA inspections and 1 compliance action at this facility.

What FDA found at this site

Latest inspection
Oct 19, 2017: Voluntary Action Indicated (VAI)
Inspections on record
6 (1 OAI, 1 VAI, 4 NAI)
Last OAI inspection
Jun 23, 2015
Warning letters
1 on record (posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once), latest Feb 26, 2016. Many letters do not state an FEI: search warning letters by company name
Form 483 and inspection documents
1 in FDA's reading room
Recalls as the recalling firm
2, latest Aug 20, 2019

A Form 483 lists an inspector's observations. OAI means FDA recommends regulatory and/or administrative action. A warning letter is FDA's formal enforcement notice. No record here does not establish that a site is safe. Check this site on FDA's Data Dashboard.

Inspections and what FDA cited

Newest first. Citations are the inspector observations FDA publishes for the inspection.

DateResultWhat FDA cited
Oct 19, 2017Voluntary Action Indicated (VAI)Lack of or inadequate procedures (21 CFR 820.100(a))
Design validation - simulated testing (21 CFR 820.30(g))
Dec 16, 2016No Action Indicated (NAI)Lack of or inadequate complaint procedures (21 CFR 820.198(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Acceptable supplier records, inadequate records (21 CFR 820.50(a)(3))
Design changes - Lack of or Inadequate Procedures (21 CFR 820.30(i))
and 4 more
Jun 23, 2015Official Action Indicated (OAI)Lack of or inadequate procedures for distribution (21 CFR 820.160(a))
Design verification - output does not meet input requirement (21 CFR 820.30(f))
Quality audits - Lack of or inadequate procedures (21 CFR 820.22)
Design plans - Lack of or inadequate (21 CFR 820.30(b))
and 12 more
Jul 20, 2012No Action Indicated (NAI)No citations listed
Sep 21, 2010No Action Indicated (NAI)No citations listed
Oct 28, 2009No Action Indicated (NAI)Documentation (21 CFR 820.75(a))
Process control procedures, Lack of or inadequate procedures (21 CFR 820.70(a))

Warning letters

Newest first. Posted FDA warning letters and Data Dashboard warning letters for this FEI, counted once. Many letters do not state an FEI, so this list can be incomplete: search warning letters by company name.

Recalls

Newest first. Recalls FDA's Enforcement Report lists with this FEI as the recalling firm. The recalling firm is not always the site that made the product. Search FDA's Enforcement Report by recall number.

DateClassRecallProduct and reason
Aug 20, 2019Class IIZ-0051-202020" 2-Needle 26-Gauge 6mm HlgH-Flo Subcutaneous Safety Needle Sets" Product # RMS22606…
Mislabeled needle lengths
Mar 10, 2016Class IIZ-0317-2018Freedom60 Precision Flow Rate Tubing Sets, REF/Part Numbers F900, F2400, F60, F10, F275… and 1 more product
On 2/24/2016 during post sterilization inspection of part number F180 Precision Flow Rate Tubing, Lot number 1.042/16 the firm found 701 out…

Form 483s and inspection documents

  • 483 on FDA.gov
    Inspection Nov 29, 2016 to Dec 16, 2016, issued Dec 16, 2016 (dates printed on the document). FDA listing date May 22, 2017. Published by FDA Jan 18, 2024.

FDA record score

3 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.

Counted on Nov 25, 2026, method version 1. It counts records and does not judge the site: a low number does not establish that a site is safe, and sites that FDA inspects more often can collect more points. How this is counted.

5 records behind this number
RecordDatePoints x age weightAdds
Inspection classified VAIOct 19, 20176 x 0.251.5
Class II recall Z-0051-2020, this FEI as the recalling firmAug 20, 20194 x 0.251
Class II recall Z-0317-2018, this FEI as the recalling firmMar 10, 201610 years or older, so it adds no points0
Warning letter on FDA's dashboard, which does not show close-outsFeb 26, 201610 years or older, so it adds no points0
Inspection classified OAIJun 23, 201510 years or older, so it adds no points0

When each source was last refreshed

FDA sources behind this page and when RegDataLab last refreshed each one
RecordsLast refreshedSource as ofStatus
InspectionsNov 25, 2026Newest record Nov 20, 2026Current
Inspection citationsNov 25, 2026Newest record Nov 17, 2026Current
Compliance actions (warning letters, injunctions and other action types)Nov 25, 2026Newest record Nov 19, 2026Current
Warning letter textNov 24, 2026Newest record Nov 17, 2026Current
Recalls (Enforcement Report)Nov 25, 2026Newest record Nov 25, 2026Current
Form 483s and inspection documentsNov 18, 2026Newest record Nov 17, 2026Current
Import refusalsNov 22, 2026Newest record Nov 12, 2026Current

Reviewing Repro-Med Systems, Inc. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.

Check this supplier

Or check Repro-Med Systems, Inc. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.

Get a free API key

Facility details

Company
Repro-Med Systems, Inc.
Location
Chester, NY (US)
Source
inspections

Sources: openFDA, FDA Data Dashboard, FDA.gov. Report a data issue: hello@regdatalab.com.