NUMED, INC. FDA Inspection Record
NUMED, INC. is an FDA-registered facility (FEI 1318694) located in Hopkinton, NY, US. RegDataLab tracks 6 FDA inspections at this facility.
What FDA found at this site
- Latest inspection
- Apr 13, 2017: No Action Indicated (NAI)
- Inspections on record
- 6 (2 VAI, 4 NAI)
- Last OAI inspection
- None on record
- Warning letters
- None matched to this FEI. Many letters do not state an FEI: search warning letters by company name
- Recalls as the recalling firm
- 2, latest Apr 21, 2025
Inspections and what FDA cited
| Date | Result | What FDA cited |
|---|---|---|
| Apr 13, 2017 | No Action Indicated (NAI) | No citations listed |
| Apr 29, 2016 | Voluntary Action Indicated (VAI) | Contamination control, Lack of or inadequate procedures (21 CFR 820.70(e)) Sampling methods - Lack of or inadequate procedures (21 CFR 820.250(b)) Documentation (21 CFR 820.100(b)) Lack of or inadequate process validation (21 CFR 820.75(a)) |
| Oct 10, 2014 | Voluntary Action Indicated (VAI) | No citations listed |
| Aug 23, 2012 | No Action Indicated (NAI) | No citations listed |
| Jun 10, 2010 | No Action Indicated (NAI) | No citations listed |
| Oct 5, 2009 | No Action Indicated (NAI) | No citations listed |
Recalls
| Date | Class | Recall | Product and reason |
|---|---|---|---|
| Apr 21, 2025 | Class II | Z-1754-2025 | BIB Balloon in Balloon Dilatation Catheter. Model Number: 420 Balloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the… |
| Oct 17, 2018 | Class II | Z-0510-2019 | Multi-Track Angiographic Catheter (MMTA)
Product Code: MMTA05100
Product… and 1 more product Instructions for Use booklets were not included on the outer pouch |
FDA record score
5 of 100. A count of this site's adverse public FDA records, weighted by how recent each one is. A higher number means more, or more recent, records. Each record below adds the points shown.
4 records behind this number
| Record | Date | Points x age weight | Adds |
|---|---|---|---|
| Class II recall Z-1754-2025, this FEI as the recalling firm | Apr 21, 2025 | 4 x 1 | 4 |
| Class II recall Z-0510-2019, this FEI as the recalling firm | Oct 17, 2018 | 4 x 0.25 | 1 |
| Inspection classified VAI | Apr 29, 2016 | 10 years or older, so it adds no points | 0 |
| Inspection classified VAI | Oct 10, 2014 | 10 years or older, so it adds no points | 0 |
When each source was last refreshed
| Records | Last refreshed | Source as of | Status |
|---|---|---|---|
| Inspections | Nov 22, 2026 | Newest record Nov 9, 2026 | Current |
| Inspection citations | Nov 22, 2026 | Newest record Nov 5, 2026 | Current |
| Compliance actions (warning letters, injunctions and other action types) | Nov 22, 2026 | Newest record Nov 13, 2026 | Current |
| Warning letter text | Nov 22, 2026 | Newest record Nov 17, 2026 | Current |
| Recalls (Enforcement Report) | Nov 22, 2026 | Newest record Nov 18, 2026 | Current |
| Form 483s and inspection documents | Nov 18, 2026 | Newest record Nov 17, 2026 | Current |
| Import refusals | Nov 22, 2026 | Newest record Nov 12, 2026 | Current |
Reviewing NUMED, INC. as a supplier? Request a free FDA snapshot: name up to five suppliers and get a PDF of their public FDA record, with sources and dates. You do not need an API connection. Each snapshot is prepared by hand.
Or check NUMED, INC. yourself in Claude, Cursor or ChatGPT: get the full FDA risk summary, with inspections, warning letters, Form 483s and import alerts by site.
Facility details
- Company
- Numed Inc
- Location
- Hopkinton, NY (US)
- Source
- device